INSTRUCTION MANUAL
7600 series Tissue Bath Coolers
for use with all
Campden Instruments Vibrating Microtomes
Campden Instruments UK
Campden Instruments USA
PO Box 8148
at Lafayette Instrument Co.
Loughborough
PO Box 5729
Leicestershire
Lafayette
LE12 7TJ
Indiana 47903
Tel: (+44) 01509 817700
Tel: (+1) 765 423 1505
Fax: (+44) 01509 81770
Fax: (+1) 765 423 4111
www.campdeninstruments.com
March 2006
DCN:
Document Ref: 7600 Eng v1.0
1

Advertisement

loading

Summary of Contents for Campden instruments 7600 Series

  • Page 1 INSTRUCTION MANUAL 7600 series Tissue Bath Coolers for use with all Campden Instruments Vibrating Microtomes Campden Instruments UK Campden Instruments USA PO Box 8148 at Lafayette Instrument Co. Loughborough PO Box 5729 Leicestershire Lafayette LE12 7TJ Indiana 47903 Tel: (+44) 01509 817700...
  • Page 2: Ec Declaration Of Conformity

    Additionally instruments will not be accepted without prior authorisation and have been certified as being uncontaminated with any material that may be hazardous to the health of service personnel. Returns Authorisation and Decontamination Certificate forms can be obtained by contacting Campden Instruments. 1. Introduction...
  • Page 3 In some countries this may be a statutory requirement. Your local Health and Safety Executive (or equivalent) will be able to advise on this matter. 1.1 Overview The 7600 Tissue Bath Cooler Unit is intended for use with the Campden Instruments Range vibrating microtomes. Unfixed brain slices sectioned at 4 C give better tissue preservation and are viable for longer in-vitro recordings.
  • Page 4 Campden Instruments tissue cutter without modification and the bath chamber has the same volume as the standard Campden Instruments tissue bath (752/2B). The tissue holder is the standard Campden Instruments tissue holder (752/2A).
  • Page 5 Electrical Inlet Coolant Connection Ports Fig. 1 Tissue bath assembly 2.2 Temperature Controller Set-up Before connecting the temperature controller to a mains supply, the unit must be set for your particular voltage supply. The voltage is set by prising out the fuse holder drawer and re-inserting it such that the voltage legend for your supply is aligned with the mark on the inlet moulding.
  • Page 6 (Campden Instruments model 7600-HTP). The bath should be then ‘wrung’ onto the cooling head to ensure a good heat transfer path and the screws fitted...
  • Page 7 and tightened evenly. Do not over-tighten the screws. The heat-conducting compound should be applied in an even layer sufficiently thick to ensure full area contact. (A thin layer of compound will transfer heat more efficiently than a thick layer). The bath may be autoclaved at 140 degrees Celsius for 30 minutes. All steels (including the so-called stainless steels) will corrode (rust) if left immersed in physiological saline/a.c.s.f./buffer solutions, the rate of corrosion will increase as the solution evaporates and the corroding concentrate increases.
  • Page 8 (Note that the actual temperatures achievable will be dependent upon the solutions used and local temperature conditions) Voltage requirements 230V 50Hz or 115V 60Hz Power Rating Inlet Fuse Rating T1.25A For further information contact: CAMPDEN INSTRUMENTS, PO BOX 8148 LOUGHBOROUGH, LEICESTERSHIRE LE12 7XT. Telephone: (+44) 0150 981 4790 Fax: (+44) 0150 981 7701 Mail@campdeninstruments.com...