Download Print this page

Unicare Health Emergency Bed Power Unit Instruction & Safety Manual

Advertisement

Quick Links

Emergency Bed Power Unit (EBPU)
Instruction & Safety Manual
Please read this manual before operation, for
maintenance instructions and safe usage
The enabling power
of applied knowledge

Advertisement

loading

Summary of Contents for Unicare Health Emergency Bed Power Unit

  • Page 1 Emergency Bed Power Unit (EBPU) Instruction & Safety Manual Please read this manual before operation, for maintenance instructions and safe usage The enabling power of applied knowledge...
  • Page 2 Emergency Bed Power Unit, Instruction & Safety Manual Installation & Operation Upon receipt of the EBPU, inspect the shipping carton for damage. If there is any obvious damage please report it immediately to your supplier. If there is no damage to the shipping carton, unpack the unit, inspect the unit for damage.
  • Page 3 SOLID TWICE PER SEC Overload, remove load SOLID CONTINUOUS UPS faulty, switch unit off UPS battery discharger, sleep mode FOR FURTHER TECHNICAL SUPPORT PLEASE CONTACT THE UNICARE HEALTH TEAM ON: 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 4: Specifications

    Emergency Bed Power Unit, Instruction & Safety Manual Specifi cations Ratings: Voltage Amperes (VA) 1000 1400 Input: Rated AC Input Voltage 230V±25% AC Input Voltage Window 165-275V Input Frequency 40-70Hz Output: AC Output Voltage (Typical) 230V AC Output Voltage Tolerance Nominal ±10%...
  • Page 5: Installation

    Installation 1. Connect the power cord to a verifi ed grounded 3-wire receptacle. 2. Turn on the power switch of EBPU, the green and yellow LED will be lit indicating AC utility voltage is present. 3. Charge the batteries for 10hours before use. You may use the EBPU immediately without recharging, but the backup time may be less than the rating.
  • Page 6 Emergency Bed Power Unit, Instruction & Safety Manual Specifi cations Batteries Nominal Voltage (VOC) No. x Capacity (Ah) 1x7 2x7 2x7 Type 12V SLA Service Life 3-5 years (depending on use) Battery Recharge 90% 8 hours Autonomy >6minutes load dependent...
  • Page 7 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
  • Page 8 PO Box 82 Bentley WA 6982 Authorised Distributor T 08 9351 7500 F 08 9350 5299 info@unicarehealth.com.au www.unicarehealth.com.au...

This manual is also suitable for: