Vescent D2-135 Quick Setup Instructions

Offset phase lock

Advertisement

Quick Links

2021/09/01 05:04
Quick Setup Instructions (Offset Phase Lock
D2-135)
Please see the full
D2-135 manual
this product.
Please refer to the
Quick Setup Instructions
Laser Controller to the Lasers.
At this point you should have two lasers temperature stabilized and running at their operating current.
The two lasers should be within a few GHz of the same frequency of each other. (If unsure of the laser
frequency, you can use a spectroscopy module to get one laser on transition and then temporarily
move the spectroscopy module to the other laser to get it near the same transition.)
D2-250 or D2-150 Alignment
This section assumes you are using the
between two lasers. If you have obtained an optical beat note by other methods, please skip this
section. The new
D2-250 Heterodyne Module
similar to that of the D2-150.
Schematic drawing of the laser alignment into the
D2-150 Heterodyne Module.
Product Manuals - https://www.vescent.com/manuals/
1/5
for more detail on the performance, specifications, and features of
for setting up the electronics modules and connecting the
D2-150 Heterodyne Module
has replaced the D2-150. Alignment of the D2-250 is
Quick Setup Instructions (Offset Phase Lock D2-135)
to obtain an optical beat note
Fig. 1:
Place the two lasers and the D2-150 or
D2-250 as shown in either
figure 2
depending on which unit you are
using: the two lasers oriented 90° relative to
each other and both pointing at the
figure 1
or

Advertisement

loading

Summary of Contents for Vescent D2-135

  • Page 1 2021/09/01 05:04 Quick Setup Instructions (Offset Phase Lock D2-135) Quick Setup Instructions (Offset Phase Lock D2-135) Please see the full D2-135 manual for more detail on the performance, specifications, and features of this product. Please refer to the Quick Setup Instructions for setting up the electronics modules and connecting the Laser Controller to the Lasers.
  • Page 2 Aim for 500 μW of power although a few hundred microwatts should be sufficient. detector in the D2-160 or D2-135-FC can be damaged by more than 1 mW input power, so be careful not to couple in too much light.
  • Page 3 D2-135.) Connect the Ramp TTL on the D2-135 to the trigger on your O-scope and view the Error In Monitor to the O-scope. Set the D2-135 to N=16, VCO Low mode, put the Servo in Ramp mode and by tweaking the laser current, you should see the Error In Monitor look something like what is shown in figure...
  • Page 4 Center the O-scope on one of shown the lock points and turn down in blue, the ramp amplitude. Flip the D2-135 Servo to Lock mode. If lock the Servo Output jumps to points are shown in red. +/-10V, repeat with the opposite gain sign.
  • Page 5 2021/09/01 05:04 Quick Setup Instructions (Offset Phase Lock D2-135) the center of the beat note, as shown figure Help For further assistance, contact us at: +1 (303) 296-6766 or info@vescent.com Please see the full D2-135 manual for more detail on the performance, specifications, and features of this product.