Contents list Cover sheet Simplex 400/450 Page Contents list, copy right and complete address Page Technical datas Page Versions and handling Page Seals and maintenance Page Maintenance Page Copyright The copyright to these Operating and Installations is held by R. Schütt GmbH.
Opening way: approx 5 mm Flow diameter: Ø 10 mm Versions: different in- and output sizes your specification Maintenance time: max. 5 sealing, time to change approx 3-5 minutes Simplex 400 Different input connections Different output connections Cleaning connection Simplex 450...
Page 4
Handling CIP cleaning completely cleans the Simplex 400 and 450. The Simplex 450 has the advantage that it can be cleaned and sterilized through the cleaning input before sampling. After each sample the Simplex 450 should be cleaned, then close the output and fill the sampler with alcohol through the cleaning input, then close the cleaning input.
Seals and maintenance AE00.0318a flat seal1/2“ AE00.0332 O-ring AE00.0330 retaining ring A00.0334 PTFE-seal cone AE01.0232 o-ring 27x2 Maintenance 1. Step Release locking screw with Phillips screwdriver 2.Step Remove handwheel cap 3. Step Remove retaining ring 4. Stepp Remove outer O-ring Page 05...
Page 6
Maintenance 5. Step Unscrew the spindle with the handwheel cap 6. Step Tapping the Teflon seal cone off the spindle with a flat-head screwdriver 7. Step Remove O-ring After replacing the seals, reassemble the sampler in reverse order. To more easily press the Teflon seal cone onto the spindle, it should be heated to 50-70°...