TESTO 0638.1445 Handling Instructions

Pressure probe

Advertisement

Quick Links

Find Quality Products Online at:
Please read before using instrument!
Observe measuring range (max. excessive pressure), too high pressure values can dest-
Please read the appropriate instruction manual before using the instrument!
In order to obtain reproducable measurements with differential pressure probes and to
maintain the accuracy ±3Pa with the 10 hPa probe, the following should be observed:
- The probe must be adjusted to the ambient temperature before the start of the
measurement!
- Zero calibration must be carried out before beginning a measurement, the position of
the pressure probe should not be changed during the measurements that follow!
- The pressure probe can be positioned as required.
- Avoid changes in temperature during the measurement!
- Observe + and - inputs!
GlobalTestSupply
www.
Handling instructions
0638.1445 (±10 hPa), 0638.1545 (±100 hPa)
roy the sensor!
The pressure probe works with a differential pressure
according to the principle of a strain gauge.
The pressure inlets in the differential pressure probe are
marked with + and -. The higher pressure should be
applied to the + inlet to achieve the best possible
accurary level. Higher pressure at the + inlet produces
a positive pressure indication, while higher pressure at
the - inlet produces a negative pressure indication.
Set the connected probe to its position of use before
beginning a measurement. We recommend the use of a
magnetic holding fixutre (see ordering data). Zero calibrati-
on should be carred out before the
measurement in order to guarantee reproducable measu-
rements. The position of the probe should not be altered
during the measurement that follows! Zero calibration is
not necessary for measurements with the absolute pres-
sure probe (0638.1645). The pressure probes are calibra-
ted in the factory, calibration is carried out when the pres-
sure probe is lying flat (normal position). We therefore
recommend that the probe remains in this position during
measurements.
Make sure that the pressure probe is not subject to vibra-
tions during the measurement (vibrations effect the results
of the measurement).
Pressure probe
0638.1645 (2 bar abs)
Description
Connecting the probes
Application
.com
sales@GlobalTestSupply.com

Advertisement

loading

Summary of Contents for TESTO 0638.1445

  • Page 1 Handling instructions Pressure probe 0638.1445 (±10 hPa), 0638.1545 (±100 hPa) 0638.1645 (2 bar abs) Please read before using instrument! Observe measuring range (max. excessive pressure), too high pressure values can dest- roy the sensor! Please read the appropriate instruction manual before using the instrument! In order to obtain reproducable measurements with differential pressure probes and to maintain the accuracy ±3Pa with the 10 hPa probe, the following should be observed:...
  • Page 2 Make sure that no liquid gets into the connec- tion nipple! Ordering data Pressure sensor ±100 Pa ......0638.1345 Pressure sensor ±10 hPa ......0638.1445 Pressure sensor ±100 hPa ......0638.1545 Pressure probe 2bar abs ......0638.1645 Magnetic holding fixture.......0554.0225 Silicone hose ..........0554.0440 Warranty Pressure probe ..........24 months...

This manual is also suitable for:

0638.15450638.1645