Planet Lighting ULED Series Installation & Maintenance Instructions Manual

Spring balanced examination light

Advertisement

Quick Links

1.0

INTRODUCTION

This installation and instruction manual describes how to install, maintain and operate the Planet Model
'ULED' examination light. Prior to installation carefully read through all the instructions enclosed in this
manual. This information is subject to change and applies to Standard ULED only. Variations may apply
to custom made fittings.
If there are any concerns or you do not understand, please contact customer service at Planet Lighting.
WARNINGS
Only qualified electricians can install the Planet model 'ULED' light. It is to be direct wired only.
Installation performed by unauthorised personnel could result in personal injury and or equipment
damage. Under no circumstances should any modifications or adjustments, other than those detailed
under section 9.0 Maintenance, be made to the Planet Model ULED.
2.0

PRODUCT INFORMATION

The Planet ULED is a state of the art examination light which utilises cutting edge Solid State Lighting
technology combined with the proven Planet Model U spring arm and bracket system. Designed and
manufactured in Australia by Planet Lighting (established 1911) an R&D driven, exporting designer and
manufacturer with over 6 years experience in the design, manufacture and supply of Solid State Lighting
devices worldwide. The Planet ULED provides an improved working environment, significantly reducing
on-going maintenance, with low operational costs and offers a "greener" alternative to similar products.
The Planet model ULED allows the accurate positioning of the lamp head above any work surface. It is
designed to be mounted on a wall or ceiling and be pulled down from above when required, remaining
there until re - positioned.
At Planet we take great pride in the design and production of our lights and we are sure you will derive
satisfaction from owning and using our products. When installed and maintained in accordance with the
instructions contained in this service leaflet, this product will provide many years of trouble free opera-
tion.
Should you have any enquiry, comment or special lighting need we would like to hear from you. We back
our products with pride, service and a five year warranty.
2.1.0

Environmental Considerations

2.1.1
Standard Packaging
I box containing a complete ULED lamp with Installation and users manual.
2.1.2
Transport
transport to be carried out by road, while maintaining the following environmental
conditions.
Temperature: -15/+60 degrees Celsius
Humidity : 10/75%
Atmospheric pressure(h/Pa) 500/1060
2.1.3
Storage
The devices packaged must be stored (warehoused) in a dry place and as per the same
environmental conditions mentioned above.
2.2.0

Disposal and Recycling Information

The ULED light has been designed, taking into account what the manufacturing
components are made out of. This is to ensure when it reaches its end of life, the
disposal method with the lowest/ least negative impact, on human health and the
environment is selected. Unlike the sources of traditional light, LED's do not contain
mercury, toxic gases, filaments or fragile parts. Due to the fixture being over 95%
recyclable and there being no Bulbs/Globe changing during the ULED life of 50,000+
hours your environmental footprint will be significantly reduced.
When your ULED light reaches its end of life, contact Planet lighting or our local
agents to learn about recycling options.The ULED light must be disposed of properly
according to local laws and, regulations.
3.0

PLANNING

Prior to installing the Planet Model 'ULED' Ceiling and Wall mounts the following items should be checked
and confirmed.
3.1.0

Architectural considerations

3.1.1
Ensure appropriate mounting structure behind the wall/ceiling
3.1.2
For structural loadings and appox product weights (refer Fig12)
3.1.3
To confirm ceiling/wall height measurements at the appropriate nominal
recommended height from the floor refer Fig 1 (back page)
3.2.0

Electrical considerations

3.2.1
The Planet Model 'ULED' wall and ceiling mount are designed to be direct wired only
3.2.2
240V AC supply is available at the connection points

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ULED Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Planet Lighting ULED Series

  • Page 1 This information is subject to change and applies to Standard ULED only. Variations may apply to custom made fittings. If there are any concerns or you do not understand, please contact customer service at Planet Lighting. WARNINGS Only qualified electricians can install the Planet model ‘ULED’ light. It is to be direct wired only.
  • Page 2 INSTALLATION: The Planet Model ULED arm is designed to be mounted above the immediate work area. Standard mounting fixtures for Planet Model ULED are: Ceiling Mount & Wall Mount (specialised mountings are available on application). In all cases the suggested mounting height (measured at swivel head) is two metres from floor level. (refer Fig 1) To ensure correct balance of the Model ULED movement the lamp must be mounted with the stem, around which it rotates, vertical at all times.
  • Page 3 WARNING Take care to remove any dust during the drilling and make sure not to damage any existing cabling 5.4.0 Remove cover from the wall mount (ref Fig 2) , feed mains cable through the larger bottom hole , place correct wall mount into position and attach to wall using appropriate fixture devices 5.5.0 Electrical Connections 5.5.1 Connect mains cable to appropriate locations on terminal block (ref Fig 2)
  • Page 4 CEILING MOUNT: WARNING Ensure that before commencing the installation, the MAINS power circuit is isolated (off ) CHECK AND ENSURE THE FOLLOWING BEFORE COMMENCING INSTALLATION: • You have the appropriate mounting template • You have appropriate tools for the job • Confirm mains cable entry is located appropriately • The mounting structure is suitable • The structural loadings and appox product weights (refer Fig12) •...
  • Page 5 Fig: 7 Fig: 6 NOTE: This section only applies if continuous rotation attachment is fitted UTW - TWIN CEILING MOUNT Fig: 8...
  • Page 6 MAINTENANCE: Planet Lighting recommend that if there is a technical fault with one of our ULED lights, the product is returned to our factory to be repaired. If a qualified technician is available to repair the fixture, circuit diagrams, parts lists, and descriptions will be made available to assist in this process.
  • Page 7 Fig: 10 Depress button to remove or replace light head Fig: 11 Spring tension nut left hand (heavy) spring tube Lock Nut WARNING UNDER NO CIRCUMSTANCES SHOULD THE SPRING TENSION NUTS (TWO) BE COMPLETELY UNDONE...
  • Page 8 Fig: 12 Static weights of complete fixture (Mount, Arm & Light Head) Wall Mount 5.5kg Ceiling Mount 6.0kg RELATIVE VERTICAL MASS: 2.7kg POSITION SHOWN HERE (ARM+HEAD ON L Y) IS ESTIMATE ONLY. 445 NOM (= ~12 Nm) 610 NOM (= ~16 Nm) Note: For the purposes of quoting minimum structural requirements of mountings, torque-load of 12 Newton Metres plus lever-arm effect of length of ceiling-mount tube or wall mounting must be...

This manual is also suitable for:

Uled-wallUled-ceiling

Table of Contents