MEASURING CABLE REPLACEMENT
INSTRUCTIONS FOR
PT1 SERIES
For Models
PT1A, PT1DC, & PT1MA
Celesco Transducer Products 7800 Deering Avenue, Canoga Park, CA 91309 Tel: 800-423-
5483, Fax 818-340-1175
Created 4/19/00 C:data/PT1man
Page 1 of 4

Advertisement

loading

Summary of Contents for celesco PT1 Series

  • Page 1 MEASURING CABLE REPLACEMENT INSTRUCTIONS FOR PT1 SERIES For Models PT1A, PT1DC, & PT1MA Celesco Transducer Products 7800 Deering Avenue, Canoga Park, CA 91309 Tel: 800-423- 5483, Fax 818-340-1175 Created 4/19/00 C:data/PT1man Page 1 of 4...
  • Page 2 If rough spots, resistance, or “soft” stops are felt, the potentiometer is damaged and must be replaced. Contact Celesco for new potentiometers and replace if necessary. Extend the measurement cable a short distance and cut it with a pair of wire cutters.
  • Page 3 If rough spots, resistance, or “soft” stops are felt, the potentiometer is damaged and must be replaced. Contact Celesco for new potentiometers and replace if necessary. Remove the pot side plate and set aside.
  • Page 4 Replace the pot side plate and, without replacing the cover, temporarily screw in the socket head cap screws that hold the cover on. Make sure the bearings are seated correctly. Replace the spring side plate. Make sure the bearings are seated correctly. Replace the phenolic disk against the spring side plate.