Download Print this page

Gira Plus 5981 00 Short Instructions

Nurse call system, power rectifier for installation on din top-hat rail/for wall installation

Advertisement

Quick Links

GB
Gira nurse call system Plus
Short instructions
Power rectifier Plus 5981 00 for installation on DIN top-hat rail
Power rectifier Plus 5998 00 for wall installation
Gira
Giersiepen GmbH & Co. KG
Electrical installation
systems
P.O. Box 1220
42461 Radevormwald
Phone +49 (0) 2195 / 602 - 0
Fax
+49 (0) 2195 / 602 - 191
www.gira.com
info@gira.com
Safety instructions
Installation and mounting of electrical devices may only be
carried out by qualified electricians.
Failure to observe the instructions can result in damage to the
device, fire or other dangers.
Isolate before working on the device or load. Take account of all
circuit breakers supplying dangerous voltage to the device or
load.
i
Provide circuit breaker!
Connect a circuit breaker type D, max. 16 A upstream from the
power supply unit.
Function
Power supply unit with input voltage of 230 V AC +/-15 %.
Output voltage 24 V DC.
The device is designed for 24-hour operation at rated output.
i
Note: read the system operating instructions.
Become familiar with all the possibilities of the device and the
complete call system. Please read the
system operating instruc-
available in various languages on the CD-ROM enclosed
tions
with the ward control centre.
Scope of delivery
Order No.: 5981 00, power rectifier Plus with carrier for installa-
tion on DIN top-hat rail.
or
Order No.: 5998 00, power rectifier Plus for wall installation.
Installation
The power rectifier Plus must be installed in a dry, dust-free and well
ventilated location.
Order No.: 5981 00 intended for installation on top-hat rail.
Order No.: 5998 00 intended for wall installation.
Important! Do not install in patient's rooms!
The device must be securely installed. Access to the device for
skilled electrical personnel must be possible at all times.
Important! Danger of overheating!
With an unfavourable installation location, heat accumulation may
cause device overheating. This may cause damage. Ensure sufficient
air circulation to prevent heat accumulation.
Device overview
The power rectifier 5998 00 has an LED status display in the device
cover:
The LED lights up green when mains voltage is applied.
Legend:
Terminals for input voltage 230 V AC
Terminals for output voltage 24 V DC
Micro-fuses: input T 2 A, output T 6.3 A
Status LED: green when mains voltage applied.
Fig.
1: The PCB of the power rectifier

Advertisement

loading

Summary of Contents for Gira Plus 5981 00

  • Page 1 Provide circuit breaker! The LED lights up green when mains voltage is applied. Power rectifier Plus 5981 00 for installation on DIN top-hat rail Connect a circuit breaker type D, max. 16 A upstream from the Power rectifier Plus 5998 00 for wall installation power supply unit.
  • Page 2 (specialist trade/installation company/electrical specialist trade). Mains frequency: 45 to 65 Hz They will forward the devices to the Gira Service Center. Output (secondary voltage) Output voltage in mains operation: 27.2 V (+/- 0.5 %)

This manual is also suitable for:

Plus 5998 00