Concentric Ring Probe
Installation, Operation and Maintenance
Figure 1. Charleswater
Figure 2. Charleswater
Kit
Description
The Charleswater Concentric Ring Probe is an instrument to
be used in conjunction with a resistance megohmmeter, such
as the Charleswater
99026
to measure surface resistance per IEC 61340-2-3, the test
method listed in Packaging Table 4 in EN 61340-5-1.
The Concentric Ring Probe can measure the volume
resistance of planar materials using a flat conductive metal
plate (not included).
The Concentric Ring Probe may be used for the resistance
measurements of ESD packaging including static shielding
and other bags.
PPE-5076.E Page 1 of 2
August 2013
TECHNICAL BULLETIN PPE-5076.E
99028
Concentric Ring Probe
99120
Concentric Ring Probe
Digital Surface Resistance Meter,
Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom
Phone: 00 44 (0) 1892-665313 • Internet:
The Charleswater Concentric Ring Probe is compatible with
the following Charleswater Surface Resistance Meters:
Item
Description
99026
Digital Surface Resistance Meter Kit
99027
Digital Surface Resistance Meter
99111
Analogue Surface Resistance Meter Kit
99113
Analogue Surface Resistance Meter
Resistivity and Resistance
When it is appropriate to convert a resistance obtained by
this test method to an equivalent resistivity in ohms per
square, multiply the resistance measurements obtained by
this method by ten. The conversion factor of ten is derived
from the geometry of the electrode assembly. NOTE:
Conversions to resistivity may not be valid for materials
that are laminated, plated or metallized with conductive
materials. This fact should be understood when users of this
test procedure are required to satisfy specifications given in
"ohms/square."
Per User guide EN 61340-5-2:1999 Clause 4.1.1
"Point-to-point resistance has been discussed, rather than
the surface and volume resistivity which was found in
previous standards and reports. This change has been
made to cater for non-homogenous materials, which are
becoming increasingly common in these applications, as well
as ease of measurement."
Repeatability
Per IEC 61340-2-3 clause 10 "Repeatability and
reproducibility, the repeatability of these test methods can be
assumed to be in the range of approximately one half order
of magnitude. If the average value for a series of laboratory
tests is 5 x 10
ohms, the spread of values can be expected
10
from 2.5 x 10
ohms to 7.5 x 10
10
Packaging
99028
Concentric Ring Probe
1 Concentric Ring Probe
1 BNC / Banana Jack Adapter
1 Certificate of Calibration
99120
Concentric Ring Probe Kit
1 Concentric Ring Probe
1 Carrying Case
1 BNC / Banana Jack Adapter
2 Banana Plug Test Leads
1 Certificate of Calibration
Charleswater.co.uk
Made in the
United Kingdom
ohms."
10
© 2013 Charleswater

Advertisement

loading

Summary of Contents for CharlesWater 99026

  • Page 1 Description 99028 Concentric Ring Probe 1 Concentric Ring Probe The Charleswater Concentric Ring Probe is an instrument to 1 BNC / Banana Jack Adapter be used in conjunction with a resistance megohmmeter, such 1 Certificate of Calibration as the Charleswater...
  • Page 2: Specifications

    Limit of Liability In no event will Charleswater or any seller be responsible or liable for any inury, loss or damage, direct or consequential, arising out of the use of or the inability to use the product.

This manual is also suitable for:

990279911199113