For example on the DTX-40, if the CCW reading is 50.30 and CW
reading is 50.50
Step 1 Determine actual deviation. (50.50 – 50.30) ÷ 2 = 0.10
(which is smaller than the allowable tolerance of 0.25).
Step 2 Calculate target calibration value, which is capacity less
actual deviation calculated above. 50.00 – 0.10 = 49.90.
Step 3 Determine calibration direction (only one direction calibra-
tion is required). Take the smaller number's direction,
which in this case is CCW (50.30).
Step 4 Zero the display. Rehang the calibration weight (2.5 kg
only for DTX-40) and calibrate in the CCW direction.
Making Adjustments
1. Remove the 2 screws in the front
cover and the warranty seal.*
2. Gently pull the front cover forward
to expose the PC board.
Note: The PC board is mounted on
the front cover and cables are con-
nected to it. Do not force.
3. Using a small flathead screwdriver adjust the VR2 pot adjust-
ment screw to the appropriate setting (screw faces right).
D D o o n n o o t t a a d d j j u u s s t t V V R R 1 1 (screw faces top).
4. After adjustment, slide the PC board back and replace the screws.
*Note: Removing the warranty seal automatically voids the warranty.
Specifications subject to change without notice.
VR2 POT
SCREWS
VR1 POT
DO NOT ADJUST
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: info@abqindustrial.net
CALIBRATION MANUAL
Cap Torque Tester
Models:
DTX and DTX2

Advertisement

loading

Summary of Contents for Imada DTX-15

  • Page 1 For example on the DTX-40, if the CCW reading is 50.30 and CW reading is 50.50 Step 1 Determine actual deviation. (50.50 – 50.30) ÷ 2 = 0.10 (which is smaller than the allowable tolerance of 0.25). Step 2 Calculate target calibration value, which is capacity less actual deviation calculated above.
  • Page 2 Conditions: Digital Cap Torque Tester Specification Table Temperature: 72° (±15°F) MODEL DTX/DTX2-15 DTX/DTX2-40 DTX/DTX2-85 CALIBRATION PROCEDURE Humidity: 10 - 60% CAPACITY (kgf-cm) 20.00 50.00 100.0 ALLOWABLE RANGE (kgf-cm) 19.90~20.10 49.75~50.25 99.5~100.5 List of Equipment ALLOWABLE TOLERANCE (kgf-cm) 0.10 0.25 Calibration arm (20 cm radius) *WEIGHT (kg) Small Screwdriver *Weight is appropriate for 20 cm radius calibration arm...

This manual is also suitable for:

Dtx-85Dtx-40Dtx2-15Dtx2-40Dtx2-85