Reference guide
The AMBIENT HIPVAC 50 COBLATION Wand provides fast, efficient, and precise ablation and resection of soft
tissue to meet the demands of the hip arthroscopy surgeon. The rigid shaft, 50 degree tip, and electrode design
provide advantages in bulk debridement and capsular resection. The device also provides accurate real time
temperature monitoring of the intra-articular fluid.
• The AMBIENT HIPVAC 50 Wand is designed for up to 15 minutes of cumulative ablation time
• Backup devices should always be available to avoid any delay prior to or during surgery

Surgical technique tips

Ensure suction is connected before using Wand
Recommended vacuum range is between 200-400mmHg
Ablation set points 7-9 are recommended
When inserting and removing the Wand, use caution and avoid contacting the distal portion of cannula and/or slotted cannula
with the shaft or electrode face as this may lead to potential electrode/shaft damage, patient injury and/or patient harm
The Wand should be activated off of the target tissue and then moved into light contact with the targeted tissue. While activated,
the Wand should be kept in constant motion.
Ensure tip is completely surrounded by irrigant solution when activated
To access Coag 2, depress and hold down one of the blue Coag buttons on the Controller face until "1" is shown. Release and depress
the button again. Only available with software 2.02 or higher.

Warnings and precautions

Excessive wear of the electrode may result from vigorous use against bony surfaces; that may lead to compromised performance
resulting in patient injury. We recommend checking the Wand periodically to ensure the electrodes are intact and the device is
functioning as designed. If the electrode is not intact, discontinue use and check the joint cavity by appropriate means.
Please refer to the Instructions for Use (IFU) packaged with the product for a complete list of Warnings, Precautions and Contraindications.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Ambient Hipvac 50 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Smith & Nephew Ambient Hipvac 50

  • Page 1 The device also provides accurate real time temperature monitoring of the intra-articular fluid. • The AMBIENT HIPVAC 50 Wand is designed for up to 15 minutes of cumulative ablation time • Backup devices should always be available to avoid any delay prior to or during surgery Surgical technique tips •...
  • Page 2 System setup Step 3 Step 1 Step 2 Step 4 Press the Controller power switch to the Connect power cord to Controller and Connect Foot Control to Controller. Connect COBLATION Wand to “on” (I) position. outlet. Controller. NOTE: The Controller will default to a set point of 7; the set point can be adjusted using the black mode button on the Wand handle or/ Foot Control or using the ablate set point adjustments on the front panel of the Controller.

This manual is also suitable for:

Asha4730-01

Table of Contents