Carbolite Gero Eurotherm 2216 Operating Instructions Manual

Advertisement

Operating Instructions

Temperature Controller
Including 2132 or 2116 Overtemperature Controller Instructions
section
1
2
3
4
5
See also the main manual for the oven or other product to
Eurotherm 2216
C
ONTENTS
which the controller is fitted.
page
2
3
5
6
6
MC11 - 1.05

Advertisement

loading

Summary of Contents for Carbolite Gero Eurotherm 2216

  • Page 1: Table Of Contents

    Operating Instructions Temperature Controller Eurotherm 2216 Including 2132 or 2116 Overtemperature Controller Instructions ONTENTS section page 2216 – Description 2216 – Basic Operating Instructions Altering the Power Limit 2132 & 2116 – Description 2132 & 2116 – Operation See also the main manual for the oven or other product to which the controller is fitted.
  • Page 2: 2216 - Description

    2216 – D ESCRIPTION 2216 Controller The model 2216 controller is made by Eurotherm, and is fitted by Carbolite Gero configured for immediate use. It is a digital instrument with PID control algorithms. It is supplied as a sleeve fitted controller.
  • Page 3: 2216 - Basic Operating Instructions

    NSTRUCTIONS Furnace or Oven Controls Most Carbolite Gero products are fitted with an “Instrument Switch” which cuts off power to the controller and other parts of the control circuit. See the instruction manual for the furnace or oven for the overall operating instructions.
  • Page 4 2216 Controller Altering the Setpoint With the display at “home”, showing the measured temperature, press Down  or Up  once to display the setpoint; press again or hold down to adjust it. The display returns to the measured temperature when no key is pressed for 0.5 seconds. Altering the Ramp Rate It is possible to limit the rate of heating by setting a ramp rate.
  • Page 5: Altering The Power Limit

    In many models the power limit setting depends on the supply voltage; usually the furnace or oven manual contains details: if in doubt, contact Carbolite Gero for advice. The power limit parameter does not apply to the overtemperature controller, if present.
  • Page 6: 2132 & 2116 - Description

    The model 2132 and 2116 overtemperature controllers are made by Eurotherm, and are fitted by Carbolite Gero configured for immediate use. They are digital instruments with a latching alarm, requiring no additional panel controls for its use. The instruments are similar in function, but the 2116 is physically larger.
  • Page 7 2132/2116 Overtemperature Controller Basic Operation Use Down  and Up  to alter the overtemperature setpoint. This should normally be set a little above the maximum working temperature (say 15°C above). The unit is supplied at 15°C above the furnace or oven maximum working temperature. Press Scroll ...
  • Page 8 Furnace and Oven products, please contact: Carbolite Gero Service Telephone: +44 (0)1433 624242 Fax: +44 (0)1433 624243 Email: ServiceUK@carbolite-gero.com Carbolite Gero Limited, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: info@carbolite-gero.com MC11 –...

This manual is also suitable for: