Omega AQM-101 Instruction Sheet

Iaq formaldehyde (hcho) logger

Advertisement

Quick Links

AQM-101
IAQ Formaldehyde
(HCHO) Logger
INSTRUCTION
M5443/1114
SHEET
Shop online at: www.omega.com e-mail: info@omega.com
For latest product manuals: www.omegamanual.info
MADE IN TAIWAN
FEATURES
1) Super large LCD simultaneously display of HCHO level,
Temp., Relative humidity, Calendar (Y/M/D) and Time
(clock).
2) Three (3) smiley icons indicate indoor air quality levels
(0/0.08/0.8ppm), easy to understand HCHO concentration.
3) Stable Electrochemical type sensor for HCHO detection.
4) High alarm threshold is selectable.
5) Manual 0ppm calibration.
6) Audible/Visible LED warning for high HCHO concentration
MAINTENANCE
Cleaning and storage:
1). The meter should be cleaned with a damp cloth and mild
detergent when necessary.
2). Store the meter in an area with moderate temperature
and humidity.
MANUAL CALIBRATION TO ZERO
Before start HCHO manual calibration make sure the area is
ventilation and no HCHO environment. Ensure the calibration
completed, to use adaptor during calibration is recommended.
HCHO meter is completed calibration in factory before
packed. Manual calibration is only for meter inaccuracy.
When power on, long press
UNIT
button more
than 6 seconds, the value will turn into "00000"
and flash
60
seconds.
After calibration completed, display will show
full LCD icon for a second. Meter will restart
automatically to return normal measurement.
BATTERY INDICATION
When battery is low (HCHO consumes power ),the reading
of HCHO will be terribly high, could be couple thousand in
ppm. "LO BAT" is flashing at the bottom of the screen,
near calendar and time. Replace new battery when LO BAT
shows.
SETTING DATE & TIME
a) Real time and date:
The meter shows Mo./Date and Hr/Min on the 3rd layer of
display, and each cycle is 16 seconds.
b) Month/Date and Hour/Min
Press
MODE+POWER
button to enter the real date and
time setting. The time default is 24-hours format.
4
INTRODUCTION
Thank you for purchasing desktop HCHO meter.
It is used to measure HCHO concentration,air temperature
and relative humidity with visible and audible alarms.
This HCHO meter is an ideal instrument for indoor air quality
(IAQ) diagnosis and HVAC system performance verification.
MATERIAL SUPPLIED
(1) Meter
(2) Universal Adaptor
(3) Operation manual
(4) 4 pcs batteries.
(5) USB cable
(6) White Box
POWER SUPPLY
The meter is powered by Super heavy duty batteries AAA x 4 pcs
(Recommended) or an AC adaptor(9V/0.5 A output)
LCD DISPLAY
Symbols:
:
1)ppm
HCHO unit
: 0ppm ~ 0.08ppm
2) icon
3) icon
: 0.08ppm ~ 0.8ppm
:
0.8ppm
4) icon
5)Air Temp.: Ambient Temperature
6)MD : Month / Date
7)HM: Hour / Minute
8)TWA: Time Weighted Average
(8 hours)
9)STEL: Short-Term Exposure Limit
(15 minutes weighted average )
10)Logging: When it is in the recording mode
11)VOC, NO2, WBGT, hpa, mbar ,EXT TEMP,lps%,cfm/p are not
available in these models
1
SETTING DATE & TIME
The meter interchangely shows current date and time
on the 3rd layer of display, and each cycle is
6 seconds.
The time default is 24-hours format.
MODE POWER
+
-->Enter date/time mode
SEL/R
--> Number goes up.
TRH/M
--> Number goes down.
MODE
-->Edit Month/date or Hour/Minute.
UNIT
-->Select to edit Month or date, Hour or Minute.
After set up, press
MODE POWER
+
to save and exit.
ALARM SETTING
Set up alarm function:
In normal measurement, press
SEL/R+POWER
bell will appear alarm function. Once set up, the red
LED will flash if the data exceeding the setting data.
HCHO Hi alarm setting.
SEL/R POWER- ->
+
Enter HCHO alarm setting.
- ->
UNIT
Change digit.
SEL/R
- ->
Number goes up.
M
D
TRH/M
- ->
Number goes down.
After set up
SEL/R POWER
+
to save and exit setting.
HCHO alarm ON/OFF:
In normal function, short press
POWER
alarm on, short press
POWER
again to turn off.
ppm
HCHO alarm on
Humidity
Humidity
o
AIR
AIR
TEMP
TEMP
M
D
5
KEYPAD (CONTROLS)
1) MODE
: HCHO mode,MAX,TWA,STEL
2) SEL./R : Edit alarm function
Logger mode: Long press to key start/off.
3) TRH.M. : Temp. Humidity mode. Setting/ Next
4) UNIT
: Select Temp. unit
5) POWER: Turn meter on/off, alarm on/off.
OPERATION
Warning: Remove batteries when not in use to prevent battery
leakage (Out of warranty) and power consumption.
(a) POWER ON/OFF
Put battery or plug in adaptor, meter turns on automatically .
LCD shows current HCHO, RH & Temperature, Date and Time.
Three(3) smiley icons indicate the indoor air quality level and
appear on the top of first layer display.
Note:
(1) When low battery, meter will restart and logger will stop if
user switch to adaptor.
(2) Please use properly power source, input voltage:
100~240 VAC,50-60Hz
ppm
Output voltage: DC 7.5~9.0V
Output current: 0.5A,or the meter will be damaged.
(3) Battery life approx. 120 hours when continuous used
(b) TAKING MEASUREMENT
Humidity
The meter starts the measurements when powered
o
on and response time is less than 30 sec.
AIR
TEMP
Note:
M
D
Do not hold the meter close to any other source of HCHO.
(c) HCHO MODE
In the HCHO Mode, press
display. The top layer of display will cycle from HCHO,
Maximum, TWA, STEL.
TEMP. HUMIDITY OFFSET
Make sure the meter is off, press
then plug in the adaptor at the same time to enter offsetting
mode.
WARNING: DO NOT USE THIS FUNCTION WITHOUT
STANDARD INSTRUMENT.
Example: Standard instrument is 30.2°C,
meter shows 30.0°C.Enter temp. offset mode, press
TRH/M
till 0.2 appear.
Offset range of Temp.and RH as below:
Temp.:-9.9°C to +9.9°C / -9.9°F to +9.9°F
RH%:-9.9% to +9.9%
a) Humidity offset:
Increase value
Decrease value
buttons ,
b) Air temp. offset
Increase value
Decrease value
After setting up short press
,
icon will flash a second. Then replug adaptor to return normal
measurement.
Logger setup:
When using datalogger function, please use batteries and adaptor
at the same time to avoid power outages and data loss.
A. Install logger driver for Windows
bell icon appear
The PL2303_Prolific_DriverInstaller_v110.exedriver is for
Windows operation system, follow the installation steps
hereunder:
ppm
1.Put the CD into CD driver, click
HCHO alarm off
PL2303_Prolific_DriverInstaller_v110.exe for installation the
o
driver.
2.The set up status.
M
D
3.Click "NEXT" to continue.
4.After complete the installation, press "Finished".
ppm
Note 1:
1)
MAXIMUM:
The unit automatically records maximum reading
and minimum reading after powered on.(MINIMUM is
always 0.00ppm for HCHO meter)
2) TWA: ( Time Weighted Average 8 hours)
The meter keeps updating the reading every minute. If
the meter has been powered on for less than 8 hours,
the TWA value will be the weighted average of readings
since powered on.
3)STEL: (Short-Term Exposure Limit in 15 minutes weighted
average)The meter keeps updating readings every
minute. If the meter has been powered on for less than
15 minutes, the STEL value shows the weighted average
of reading taken since powered on.
4)RESET MAX:
Press
MODE
switch TWA or STEL.Under MAX mode, Long
press
MODE
(d) TRH MODE : Temperature / Relative Humidity mode
In the
TRH
related humidity and air Temperature
simultaneously .
(e) Temperature unit (°C or °F) selection
Press
UNIT
temperature unit (°C or°F).
MODE
button to switch HCHO
2
B. Install the logger software for Windows XP
SEL/R+TRH/M
button
(win7/win8/VISTA)
Please follow the following steps to install the software
1. Put CD into CD driver, click "Setup.exe" to install
2.Select the destination folder you are go to save by click
"
"
Browse button
3.Once you finish selecting ,press Next
4.Select "I accept the License Agreemen:(s)" and press NEXT
button.
5.Click NEXT button when you see the following.
6.Press FINISH button to complete the installation.
Humidity
7.Restart the Windows XP system
o
AIR
TEMP
C Run the software from Windows XP
.
H
M
1. Plug the USB cable into USB port, since you have already
Installed the USB driver, computer will automatically detect
logger ,
POWER
to save setting and
"%"
2. Find "Datalogger" in START
3.Select logger software folder.
Reminder:
Software is installed completed,"Get Datalog's Identifier"shows
next page(pic.1), then press SETTING button to enter Log setting
Note: If logger is not plugged in USB port, the warning message
"Reminder: Data Logger is not plugged in USB port!" shows next
page(pic.2) Make sure logger is plugged in and choose the
correct COM port.
6
ppm
ppm
ppm
MAX
TWA
STEL
button to get MAX reading and
button to clear all the values.
Mode, the meter displays both
HUMIDITY
+ TEMPERATURE
HUMIDITY
o
AIR
TEMP
button to toggle the
HUMIDITY
HUMIDITY
o
o
AIR
AIR
TEMP
TEMP
3
"
".
PROGRAMS
(Pic.1)
7

Advertisement

loading

Summary of Contents for Omega AQM-101

  • Page 1 LCD DISPLAY 15 minutes, the STEL value shows the weighted average SHEET user switch to adaptor. Shop online at: www.omega.com e-mail: info@omega.com of reading taken since powered on. Symbols: (2) Please use properly power source, input voltage: For latest product manuals: www.omegamanual.info...