Union (2002/96/EC) and belongs to Category 9 (Monitoring and Control Instruments). You can return the instrument to us free of charge for proper environment friendly disposal. You can obtain further information from your local Narda Sales Partner or by visiting our website at www.narda-sts.it .
Page 5
• Ground connections must not be interrupted intentionally; • To prevent the possible danger of electrocution, do not remove any covers, panels or guards installed on the device, and refer only to NARDA Service Centers if maintenance should be necessary;...
Page 6
This product complies with the requirements of the Low Voltage Directive 2006/95/CE and the EMC Directive 2004/108/CE. The PMM 6000N Wideband RF Power Amplifier MUST BE USED IN THE CONFINES OF A SHIELDED ROOM, in order to comply with the requirements of the referenced EMC Directive.
• an accessories check list to verify all accessories enclosed in the packaging. PMM 6000N is a wideband RF power amplifier covering a 9 kHz ÷ 230 1.2 Introduction MHz frequency range with output power level of 10 Watts into a 50 Ohm load.
Table 1-1 lists the PMM 6000N performance specifications. The following conditions apply to the specification: • The PMM 6000N needs at least 5 minutes warm-up before operate; • The ambient temperature shall be 10° to 40° C. TABLE 1-1 Technical specifications 9 kHz ÷...
1.6 Typical Performances Typical Gain at Nominal Power [dB] [dB] 0.001 0.01 1000 [MHz] Typical Gain at Nominal Power/2 [dB] [dB] 0.001 0.01 1000 [MHz] General Information...
Page 10
Power output at 1dBc [W] 20.0 18.0 16.0 14.0 12.0 10.0 0.001 0.01 1000 [MHz] Compression at Nominal Power [dBc] [dB] 0.001 0.01 1000 [MHz] General Information...
Page 11
2nd Harmonic at Pn [dBc] 3rd Harmonic at Pn [dBc] [dBc] 0.001 0.01 1000 [MHz] General Information...
• Verify that the fuse rating is appropriate for the line voltage used. Before plugging PMM 6000N into the main supply line ensure that the line voltage is in the range specified, and that the appropriate fuse have been selected.
To clean the equipment do not use any solvent, thinner, turpentine, acid, acetone or similar matter to avoid damage to external enclosure. PMM 6000N is delivered ready to use. Remove the amplifier from his 2.12 Hardware cardboard box and connect power cable to rear panel mains plug. Push Installation the main switch to "ON"...
3 - Operating Instruction 3.1 Cooling It is necessary to ensure that the PMM 6000N forced-air cooling system be allowed to operate at maximum efficiency. Cooling depends upon the free flow of cooling air through the bottom and the rear of the unit. Take care to keep these areas free of any objects that could obstruct this free air flow.
Page 16
This page intentionally left blank Operating Instruction...
Page 17
Moreover, we are continuously improving our quality, but we know this is a never ending process. We would be glad if our present efforts are pleasing you. Should one of your pieces of NARDA equipment need servicing you can help us serve you more effectively filling out this card and enclosing it with the product.