13. Specification
Maximum Operating Life
Sensor Range
Alarm Setpoints
Calibration
Shelf Life
Operating Temperature
Humidity
Audible Alarm
Visual Alarm
16
24 months from activation, assuming 3-5 minutes of
alarm per day under normal operating conditions*
CO
0 to 1000 ppm (display: 0-200 ppm)
H
S
0 to 100 ppm (display: 0-100 ppm
2
O
0 to 30% v/v (display: 0-25% v/v)
2
Level 1
CO
35 ppm
H
S
10 ppm
2
O
23.5%
2
CO/H
S
Zero adjustment (optional Span)
2
O
Span adjustment
2
12 months: CO/H
S
2
6 months: O
2
-4°F to +122°F (-20°C to +50°C)
5-95% RH (non-condensing)
95db at 4" (95db at 10cm)
High intensity red LEDs
Level 2
100 ppm
15 ppm
19.5%
XT Series Personal Gas Detector