IBM 7220 Instruction Manual page 138

Dsp lock-in amplifier
Table of Contents

Advertisement

Appendix E, ALPHABETICAL LISTING OF COMMANDS
RS [n
RT[.]
SAMPLE [n]
E-14
[n
]]
Set/read RS232 interface parameters
1
2
The values of n
set the baud rate of the RS232 interface according to the following
1
table:
n
Baud rate (bits per second)
1
0
75
1
110
2
134.5
3
150
4
300
5
600
6
1200
7
1800
8
2000
9
2400
10
4800
11
9600
12
19200
The lowest five bits in n
following table:
bit number
bit negated
0
data + parity = 8 bits
1
no parity bit
2
even parity
3
echo disabled
4
prompt disabled
Ratio output
In integer mode the RT command reports a number equivalent to 1000 × X / ADC1
where X is the value that would be returned by the X command and ADC1 is the
value that would be returned by the ADC1 command.
In floating point mode the RT. command reports a number equivalent to X / ADC 1.
Main analog to digital converter sample rate control
The sampling rate of the main analog to digital converter, which is nominally
166 kHz, may be adjusted from this value to avoid problems caused by the aliasing
of interfering signals into the output passband.
n may be set to 0, 1, 2 or 3, corresponding to four different sampling rates (not
specified) near 166 kHz.
control the other RS232 parameters according to the
2
bit asserted
data + parity = 9 bits
1 parity bit
odd parity
echo enabled
prompt enabled