Unsecuring the instrument from the remote interface

& CAL:SEC:CODE OFF,
& CAL:SEC:CODE ON, ATU3606B

Changing the calibration security code

& CAL:SEC:CODE OFF,
& CAL:SEC:CODE ABC1234
& CAL:SEC:CODE ON, ABC1234
U3606B Service Guide
Send the CALibration:SECurity:CODE <mode>, <code>
command from the remote interface.
ATU3606B
The security code can only be changed from the remote
interface after the instrument has been unsecured.
Send the CALibration:SECurity:CODE <new_code>
command from the remote interface after the instrument has
been unsecured.
ATU3606B
See the CALIbration subsystem in the U3606B
Programmer's Reference for more information on the
CALibration:SECurity:CODE command.
Calibration Procedures
Unsecuring the Instrument for Calibration
This command unsecures the instrument for
calibration. The default security code is
ATU3606B. The security code can only be
changed when the instrument is unsecured.
This command secures the instrument from
accidental or unauthorized adjustments.
This command unsecures the instrument for
calibration. The default security code is
ATU3606B. The security code can only be
changed when the instrument is unsecured.
This command changes the security code to
"ABC1234".
This command secures the instrument from
accidental or unauthorized adjustments.