Single Zone Gas Monitor

4.5. Fuses

The SZ monitor is protected from electrical damage by two, 1 A, 250 V, type "F" fuses (page 11).
Carefully remove the fuses from their fuse clips and visually inspect each fuse for damage. Replace the
fuses as required (P/N 0004-2620).

4.6. Sample Pump

The Sample Pump (page 11) draws the gas sample into the monitor, through the IR detector, and
discharges the sample out the monitor's exhaust port. When replacing the pump, remove its inlet and
outlet tubing, disconnect the AC power wires from the pump itself, and remove pump from monitor. Install
a new sample pump (P/N 3015-5176).

4.7. End-of-Line Water Stop (Optional)

The end-of-line water stops (P/N 3015-5512) prevent contaminates and moisture from entering the
monitor. This filter should be checked regularly and replaced when signs of contamination are obvious.

4.8. Water Trap (Optional)

The water trap removes moisture that enters tubing before allowing it to enter the internal components of
the SZ unit. The water trap should be emptied often as condensate accumulates. The water trap filter
should be periodically checked and replaced when there are obvious signs of contamination.

4.9. Clock Battery

The Clock Battery (page 11) maintains the correct date and time when AC power is not applied to the
monitor. Replace this battery about every 5 years (P/N 0204-0020).
P/N: 3015-4256 Rev 13
29