Download Print this page

General Specifications; Factory Setting; Model And Suffix Codes; Optional Specifications - YOKOGAWA YTA110/KU2 User Manual

Yta series temperature transmitters (hardware)

Advertisement

Factory setting ( )

Tag No.
Input sensor type
Lower bound of calibration range
Upper bound of calibration range
Unit of calibration range
Damping constant
Sensor burnout
Output when transmitter fails
*1: When option code C1 is specified, Low takes effect (–2.5%, 3.6mADC).
*2: When option code C1 is specified, Low takes effect (–5%, 3.2mADC or less).

7.2 Model and Suffix Codes

Model
Basic Specification Codes
YTA110
. . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . .
YTA310
. . . . . . . . . . . . . . . . . . . . . .
YTA320
Output
–D . . . . . . . . . . . . . . . . . . . .
signal
–E . . . . . . . . . . . . . . . . . . . .
–F . . . . . . . . . . . . . . . . . . . .
A . . . . . . . . . . . . . . . . .
Electrical
0 . . . . . . . . . . . . . . .
connection
2 . . . . . . . . . . . . . . .
3 . . . . . . . . . . . . . . .
4 . . . . . . . . . . . . . . .
D . . . . . . . . . . . .
Built-in indicator
N . . . . . . . . . . . .
Mounting bracket
B . . . . . . . . .
D . . . . . . . . .
N . . . . . . . . .
Additional specifications
*1: Applicable for YTA320 only.
*2: Use bolts for wall mounting.

7.3 Optional Specifications

Item
Power supply voltage: 10.5 to 32 V DC
Lightning protector
Allowable current: Max. 6000A(1 40 s),
repeating 1000A(1 40 s) 100 times
Coating change
Epoxy resin coating
Painting
Color change
Amplifier cover only
Amplifier and Terminal covers
*1
Stainless Steel Housing
Housing Material: SCS14A Stainless steel
Calibration Unit
Addition of Degree F and Degree R unit
Output signal low-side in
Output signal low-side: –5 %, 3.2 mA DC or less.
*2
Transmitter failure
Sensor burnout is also set to 'Low': –2.5 %, 3.6 mA DC
NAMUR NE43
Output signal limits:
*2
compliant
3.8 mA to 20.5 mA
*2
Data Configuration
Description into "Descriptor" parameter of HART protocol.
(max. 16 characters)
*2
Sensor matching function
RTD Sensor matching function
*1 : Not applicable with other option codes, except for A, C1, D2 and CM1.
*2 : Not applicable for output signal code F.
Left blank if not specified in order
"Pt100, 3-wire" if not specified in order
"0" if not specified in order
"100" if not specified in order
"°C" if not specified in order
2 seconds
High side (110%, 21.6 mA DC)
High side (110%, 21.6 mA DC)
Temperature transmitter (1 input type)
High precision temperature transmitter (1 input type)
High precision temperature transmitter (2 input type)
4 to 20mA DC output, BRAIN communication type
4 to 20mA DC output, HART communication type
FOUNDATION Fieldbus communication type
Always A
G1/2 female
1/2 NPT female
Pg13.5 female
M20 female
Digital indicator
None
SUS304 Stainless steel 2-inch horizontal pipe mounting
SUS304 Stainless steel 2-inch vertical pipe mounting
None
/
Additional specifications
Descriptions
Munsell renotation code: NI1.5 Black
Munsell renotation code: 7.5BG4/1.5, Jade green
Metallic silver
Munsell renotation code: 7.5R4/14 Red
Failure alarm down-scale: output status at CPU
failure and hardware error is –5%, 3.2 mA or less.
Sensor burnout is also set to 'Low': –2.5%, 3.6 mA
Failure alarm up-scale: output status at CPU failure
and hardware error is 110%, 21.6 mA or more.
Sensor burnout is also set to 'High': 110%, 21.6 mA
*1
*2
T0705.EPS
Description
*1
*2
7-5

7. GENERAL SPECIFICATIONS

*2
T0703.EPS
Code
A
X1
P1
P2
P7
PR
E1
D2
C1
C2
C3
CA
CM1
T0704.EPS
IM 01C50B01-01E

Advertisement

loading

This manual is also suitable for:

Yta310/ku2Yta320/ku2