Circuit Description - Teledyne 300P Operating Instructions Manual

Percent oxygen transmitter
Table of Contents

Advertisement

2 2 2 2 2 Operational Theory
sion membrane, has a surface area of 2.48 cm
sensor block through an inlet tube between the cathode and the sensor
block cap, diffuses through the Teflon membrane, and any oxygen in the
sample gas is reduced on the surface of the cathode by the following
mechanism:
O
+ 2H
2
When the oxygen is reduced at the cathode, lead is simultaneously
oxidized at the anode by the following mechanism.
2Pb
2Pb
The electrons released at the surface of the anode flow to the cathode
surface via an external circuit. This current is proportional to the amount of
oxygen. It is measured and used to determine the oxygen concentration in
the gas mixture.
The overall reaction for the fuel cell is:
2Pb + O
The output of the fuel cell is limited by the amount of oxygen in the
cell at any one time, and the amount of stored anode material. In the
absence of oxygen, there is no current generated.
Warning: The sensor used in the Model 300P uses electrolytes which
contain substances that are extremely harmful if touched,
swallowed, or inhaled. Avoid contact with ANY fluid or powder
in or around the unit. What may appear to be plain water could
contain one of these toxic substances. In case of eye contact,
immediately flush eyes with water for at least 15 minutes. Call
physician. (See Appendix - Material Safety Data Sheet (MSDS)).
2.3

Circuit Description

The 300P circuitry consists of four stages: an input stage, a tempera-
ture compensation/ gain stage, a low-pass filter and an output voltage-to-
current stage. The output of the O
the amount of oxygen present. The input stage converts this current to a
voltage signal.
The temperature compensation stage consists of circuitry that ampli-
fies the signal by a factor which is dependent upon the O
This temperature dependent gain is used to compensate for the rise in
output current of the O
the sensor assembly monitors this temperature and changes resistance (up
2-2
– – – – –
– – – – –
O + 4e
4OH
2
– – – – –
+2
+ 4e
2PbO
2
cell is a current directly proportional to
2
cell with temperature. A thermistor mounted in
2
Teledyne Analytical Instruments
2
. The sample gas enters the
(cathode)
cell temperature.
2
Model 300P
(anode)

Advertisement

Table of Contents
loading

Table of Contents