4.1. Changing the Weighing Speed Using WinCT-AD4212C

The weighing speed can be selected from the following three rates to minimize the influence on
weighing that is caused by drafts and/or vibration at the place where the balance is installed.
Indicator
Speed
FAST
Fast
MID.
Slow
SLOW
Operation
1. Press the MODE button.
2. Press the MODE button to select a weighing speed.
Either FAST, MID. or SLOW can be selected.

4.2. Calibration

Calibration
Calibration using the calibration weight.
Caution
Do not allow vibration or drafts to affect the balance during calibration.
Caution on using an external calibration weight
The accuracy of the weight can influence the accuracy of weighing. Select an appropriate
weight as listed below. A calibration weight (conforming to OIML, Class E2 or equivalent) is
provided with the balance as a standard accessory.
Model
AD-4212C-300
AD-4212C-600
AD-4212C-3000
AD-4212C-6000
AD-4212C-301
AD-4212C-3100
The calibration weight in bold type: factory setting
Display
Stability
Sensitive value
Stable value
Usable calibration weight
50g, 100g, 200 g, 300g
50g, 100g, 200 g, 300g, 400 g, 500 g, 600 g
50g, 100g, 200 g, 300g, 400 g, 500 g, 1000 g,
2000g, 3000g
200 g, 500 g, 1000 g, 2000g, 3000g, 4000 g,
5000 g, 6000 g
50g, 100g, 200 g, 300g
50g, 100g, 200 g, 300g, 400 g, 500 g, 1000 g,
2000g, 3000g
This indicator means "the balance is measuring calibration
data". Do not allow vibration or drafts to affect the balance
while this indicator is displayed.
Weighing speed setting
MODE
16
MODE
Each pressing
switches
the indicators
Calibration weight provided
200 g