Test Lead Set With Retractable Shrouds; Esu- Dispersive Safety Lead - Fluke RF303 Operator's Manual

Electrosurgical analyzer
Table of Contents

Advertisement

RF303
Operators Manual
To avoid possible electric shock, burning of the skin, or
personal injury, follow these guidelines: Retractable end of
test leads are for use on ESU only. No probes or
accessories supplied with the Analyzer are intended for
handheld use. Set up using the safety test leads and stand
clear when activating the ESU with the footswitch.
The following is a description of the supplied test leads.

Test Lead Set with Retractable Shrouds

Part # 1903307
Color: One red and one black
Quantity: Two/set
One end of each test lead has a fixed shroud 4mm banana plug to connect with
the Analyzer. The other end has a retractable shroud 4mm banana plug
compatible with the active electrode/bipolar jacks on most ESU panels. A red
safety alligator clip is included for direct connection to an actual active
electrode element if preferred.
A removable black clamp facilitates firm attachment to a ground reference
point. This lead is important for making leakage measurements in the battery
mode of operation.

ESU- Dispersive Safety Lead

Part # 2772171
Color: Clear with blue connectors
Quantity: One
One end of the test lead has a single fixed-shroud 4mm banana plug to connect
with the Analyzer. The other end has a CQM style two-pin connector that
plugs directly into the neutral (or dispersive) electrode jack on most ESU
panels.
2-8
W X Warning

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents