Specifications Hi 9110, Hi 9210 & 9210N; Operational Guide - Hanna Instruments HI 9023C Instruction Manual

Portable waterproof ph meters
Hide thumbs Also See for HI 9023C:
Table of Contents

Advertisement

SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
& HI 9210N
& HI 9210N
& HI 9210N
& HI 9210N
& HI 9210N
HI 9110
Range
pH
°C
Resolution
pH
0.01 pH
°C
Accuracy
pH
(@20°C/68°F) °C
Typical EMC
pH
±0.05 pH
Deviation
°C
Calibration
Manual 2 points through trimmers
Offset Calibration
Slope Calibration
From 85 to 105% of nominal value
Temperature
Manual 0 to 100 C
Compensation
(32 to 212 F)
Electrode
FC 100B HI 1230B
pH (incl.) pH (incl.) pH/ C (incl.)
Input Impedance
Battery Type
4x1.5 volt, AA size (alkaline batteries);
Life
2000 hours of continuous use
Environment
0 to 50 C (32 to 122 F); 100% RH
Dimensions
196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
Weight
11
HI 9210
HI 9210N
0.00 to 14.00 pH
0.0 to 70.0 C
0.1 C
±0.02 pH
±0.5 C
±0.02 pH
±0.5 C
±1 pH
Auto 0 to 70 C
(32 to 158 F)
HI 1217D
10
12
Ohm
320 g (11.3 oz.)

OPERATIONAL GUIDE

OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with batteries.
Remove the back cover, unwrap the batteries and
install them while paying attention to polarity.
HI 9023C, HI 9024C, HI 9025C:
To prepare the instrument for use connect the pH electrode and the
temperature probe to the BNC and temperature sockets on the top of
the instrument. The temperature probe can be used independently to
take temperature measurements, or it can be used in conjunction with
the pH electrode to utilize the meter's ATC capability. If the probe is
disconnected, temperature can also be set
manually with the UP and DOWN keys. To
switch the instrument on, press and hold the
ON/OFF key for a fraction of a second. The
meter has a built-in protection against electro-
magnetic interference and the delayed response
of the keys assures that the commands are not
mistaken for stray signals.
HI 9110 and HI 9210:
Connect the pH electrode to the BNC connector on the top of the
instrument and press the ON/OFF key.
HI 9210N and HI 9023CN:
Connect the pH/ C electrode to the DIN connector on the top of
the instrument and press the ON/OFF key.