Agilent Technologies U3402A User's Manual And Service Manual page 116

51/2 digit dual display multimeter
Hide thumbs Also See for U3402A:
Table of Contents

Advertisement

6
Specifications and Characteristics
Table 6-9 Supplemental measurement specifications (continued)
Resistance (2-wire W and 4-wire W)
• Measurement method
• Open-circuit voltage
• Zeroing error
• Input protection
• Response time
Diode/continuity
• Measurement method
• Test current
• Open-circuit voltage
• Continuity threshold
• Continuity level
102
2-wire Ohms or 4-wire Ohms
Limited to < +5 VDC
0.05 W or less (excluding test lead resistance) in each range when Rel operation is used
500 V on all ranges
Approximately 1.5 seconds for 12 MW and ranges below 12 MW
Approximately 5 seconds for 40 MW
Approximately 10 seconds for 120 MW
Approximately 23 seconds for 300 MW
0.83 mA ±0.2% constant current source, open-circuit voltage limited to < 5 V
Approximately 0.5 mADC
Limited to < +5 VDC
10 W fixed
Approximately < +50 mVDC
U3402A User's and Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents