Troubleshooting Hints - EUTECH INSTRUMENTS CADMIUM EPOXY Instruction Manual

Table of Contents

Advertisement

Cadmium Electrode

TROUBLESHOOTING HINTS

Symptom
Out of Range
Reading
Noisy or Unstable
Readings (readings
continuously or
rapidly changing)
Drift (reading
slowly changing
in one direction)
Possible Causes
defective meter
defective electrode
electrodes not
plugged in properly
reference electrode
not filled
air bubble on
membrane
electrodes not
in solution
defective meter
air bubble on
membrane
electrode exposed
to interferences
defective electrode
ISA not used
meter or stirrer
not grounded
samples and standards
at different
temperatures
electrode exposed
to complexing agents
18
Instruction Manual
Next Step
check meter with
shorting strap
(see meter instruction manual)
check electrode operation
unplug electrodes
and reset
be sure reference
electrode is filled
remove bubble by
redipping electrode
put electrodes
in solution
check meter with
shorting strap
remove bubble by
re-dipping electrode
soak electrode in
cadmium standard
replace electrode
use recommended ISA
ground meter or
stirrer
allow solutions to come
to room temperature
before measurement
check section
Precipitation
entitled
and Complexation