1) Connect a level meter to terminals OUTPUT
(except for the CUE track).
2) Playback Test Tape MTT-356 and check that
the output
level stays within the specified
boundaries {see Fig. 6-3) over the entire fre-
quency range.
- azimuth again.
a
PT
| tT a
oo
WEEE
PTT
FREQUENCY
IN He
RESPONSE
0A 48
Fig. 6-3.
Playback frequency response
6-4. BIAS DUMMY COIL ADJUSTMENT
1) Connect.a.frequency counter to test point -TP2 .
at bias oscillator U401,
2) Load the deck with a blank test tape. Set all
FUNCTION
REC/MORMAL
switches at REC
and select REC/PAUSE mode with operating
buttons.
;
3) Check bias frequency (fo).
NOTE: The fo.is.a.bias frequency value deter-
mined when all the FUNCTION switches
of every channels (L, R, and CUE) are set
at REC, with the dummy coil in a non-
operating state.
4) Resetting:-4:-channel
(S103) alone to NOR-
MAL, adjust L401
until the bias frequency
becomes fo.
5) Resetting
R-channel
(S203) alone to NOR-
MAL, adjust L402 until the bias frequency
becomes fo.
6) Resetting CUE channel (S303) alone to NOR-
MAL, adjust L403
until the bias frequency
becomes fa.
7) Check
that the bias frequency stays within
+506 Hz of fo at any combination of settings
of the REC/NORMAL
switches ($103, S203,
and $303).
6-5. BIAS FILTER
ADJUSTMENT
1) Connect
a level meter to terminals OUTPUT
(L/R/CUE).
_ is —10.dBV
+1 dB and the +4 dBm
BAL
- OUTPUT. terminal level is +4-dBm +1-dB.
R249
so that the meter of the L/R-channel
read +3 VU, and semi-fixed variable resistor
R349 so that the CUE meter reads O dB.
6-7. BIAS TRAP ADJUSTMENT
6-B. BIAS ADJUSTMENT
Make
the following
preparation
for adjusting
bias.
a. Connect
measuring
instruments as shown in
Figure 6-1.
b. Load the deck with Test Tape MTT5061 and
put it in REC/PAUSE mode.
c. Set the REC/NORMAL
switch of every chan-
nel at REC,
d. Apply 400 Hz, —11 dB (282 mV) to input
terminals L, R, and CUE.
e. Set the OUTPUT control of every channel at
its maximum position.