If You Have a Problem
If you get an "operation timed-out" message
If the power to the HP 16500B was just turned ON, make sure that you
wait 15 seconds after the System Configuration menu is displayed before
attempting the TELNET.
Make sure that another user is not already accessing the file system as
the control user or connected to the command parser socket.
If you get an "operation timed-out" message
Verify the LAN connection between your computer and the HP 16500B.
Refer to "If you cannot connect to the HP 16500B" earlier in this chapter.
Increase the file time-out value on your PC or workstation.
If the HP 16500B begins to operate slowly
The HP 16500B may operate more slowly if multiple users are trying to use
the system at the same time.
Copy the file \system\mount.txt to your local computer an view it
with a text editor. Check to see whether any other users are currently
connected to the system.
If all else fails
Contact your system administrator, or consult the HP 16500H LAN
Interface Module Administrator's and Service Guide.
If you still cannot solve the problem, contact the nearest HP Service
Center for repair information.