4.6 Connecting industrial enclosure
Figure 4-6
4.6.1.1

Terminal layout (industrial enclosure)

The following table shows:
• Which cable with which terminal
• Hardware and software configuration of the channels
56
Function
Description
RL <500 Ohm (HART ≥230 Ohm)
Active: Uo=28 V, Io=87mA
Current Output
1
- 4 to 20 mA,
Passive : Ui=30 V DC, Ii=100 mA
Signal Output
RL <500 Ohm
- C: 0/4 to 20 mA
Active: Uo=28 V, Io= 87 mA
2
- F: fmax 10 kHz
Passive: Ui=30 V DC, Ii=100 mA
Active: Uo=28 V, Io=87 mA
3
Signal Input
Passive : Ui=30 V DC, Ii=100 mA
4
Relay Output
Power supply con‐
L/+
nections
N/-
Key to symbols
A
P
NO
NC
Updates (to be filled in on firmware and hardware updates)
Device configuration date
Initial firmware and hardware revisions
Configuration of channels 1, 2, 3 and 4
Configuration label
Terminal 1
Terminal 2
Terminal 3
Configured as active input/output
Configured as passive input/output
Connected as normally open contact
Connected as normally closed contact
Compact Operating Instructions, 12/2023, A5E38755607-AF