Download Print this page
Agilent Technologies E8364BH85 Service Note
Agilent Technologies E8364BH85 Service Note

Agilent Technologies E8364BH85 Service Note

Pna microwave network analyzers

Advertisement

Quick Links

M M
O O
D D
I I
F F
I I
C C
A A
T T
I I
O O
N N
M
O
D
I
F
I
C
A
T
I
O
N
S E R V I C E
E8364BH85 PNA Microwave Network Analyzers
Serial Numbers: 0000A00000/9999Z99999
Intermittent 10 GB HDD electrical problem can cause PNA HDD-related failures.
Replace 10 GB HDD with a new 40 GB HDD.
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
Part Number
E8356-60076
-OR-
E8801-60063
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
IMMEDIATELY
CATEGORY:
ON SPECIFIED FAILURE
:
X AGREEABLE TIME
LOCATION
CUSTOMER INSTALLABLE
CATEGORY:
X ON-SITE
X SERVICE CENTER
AVAILABILITY:
AUTHOR: MF
ADDITIONAL INFORMATION: This Service Note applies to the following PNA models: E8356/7/8A, N3381/2/3A, E8801/2/3A,
E8361/2/3/4A, E8362/3/4B, and E8362/3/4BH85.
© AGILENT TECHNOLOGIES, INC. 2008
PRINTED IN U.S.A.
January 16, 2008
R R
E E
C C
O O
M M
M M
E E
N N
D D
E E
D D
R
E
C
O
M
M
E
N
D
E
Description
HDD for
E8356/7/8A
HDD for:
E8356/7/8A,
E8801/2/3A,
N3381/2/3A,
E8361/2/3/4A,
E8362/3/4B
PRODUCT'S SUPPORT LIFE
D
E8364BH85-04
Operating System CPU
Windows 2000
Windows XP
STANDARDS
LABOR: 0.5 Hours
SERVICE
INVENTORY: N/A
NO CHARGE AVAILABLE UNTIL: October 31, 2008
PRODUCT LINE: WN
N O T E
Supersedes:
Qty.
266 MHz
1
500 MHz
1
USED
RETURN
PARTS:
X SCRAP
SEE TEXT
Page 1 of 2
None

Advertisement

loading

Summary of Contents for Agilent Technologies E8364BH85

  • Page 1 S E R V I C E N O T E Supersedes: None E8364BH85 PNA Microwave Network Analyzers Serial Numbers: 0000A00000/9999Z99999 Intermittent 10 GB HDD electrical problem can cause PNA HDD-related failures. Replace 10 GB HDD with a new 40 GB HDD.
  • Page 2 Page 2 of 2 E8364BH85-04 Situation: Some PNAs are experiencing reliability issues due to an intermittent electrical failure on a circuit board inside the 10 GB HDD. This can cause operating system failures to occur. These failures are eliminated by replacing the 10 GB HDD with a 40 GB HDD.