Fig.58
Fig.59
2. Friction pad
The
adjustment
sequence
is: loosen screw (a)
of arm
to be adjusted
(1 1/2 turns).
Position
the X-ray
head
in a 450angle
to the vertical
scissor
arm.
The
arm
with
the
X-ray
head
shall remain stationary at a 45° angle as shown.
When
lowering
the
X-ray
head
(to
less than
45°) the X-ray head must return to 459 angle
by
itself
when
released.
Spring
adjustment
is made with screw (d).
When
raising the X-ray
head
(to more
than
450)
the
X-ray
head
must
return
again
by
itself
when
released.
Spring
adjustment
is
made with screw (c).