2-1.
GENERAL
Before a 9000 Series Micro System Troubleshooter can be used to perform any
testing or fault isolation, it must be connected to the UUT. Connection is made
by means of the pod, which is equipped with two cable assemblies, one shielded-
type and one ribbon-type. The procedures for installing and connecting the pod
are given in the following paragraphs.
2-2.
MAKING CONNECTIONS
Before making any connections to the UUT, take note of the following
precautions:
WARNING
TO PREVENT POSSIBLE HAZARDS TO THE OPERATOR OR
DAMAGE TO THE UUT, DISCONNECT ALL HIGH-VOLTAGE
POWER SUPPLIES, THERMAL ELEMENTS, MOTORS, OR
MECHANICAL ACTUATORS WHICH ARE CONTROLLED OR
PROGRAMMED BY THE UUT MICROPROCESSOR BEFORE
CONNECTING POD.
Be sure to install the ribbon cable plug correctly m the UUT
microprocessor socket.
The self test socket is intended for use with the ribbon cable plug only. Do
not insert any microprocessor removed from a UUT under test, or any
other device into this socket.
Connect the pod between the troubleshooter and the UUT as follows:
I.
Remove power from the troubleshooter and the UUT.