Respirox
BPAP System
T-30T plus
User Manual
0123

Advertisement

Table of Contents
loading

Summary of Contents for Ece Medical Respirox T-30T plus

  • Page 1 ® Respirox BPAP System T-30T plus User Manual 0123...
  • Page 2: Table Of Contents

    BPAP System T-30T plus User Manual V2.11 Table of Contents 1. Symbols ·································································································· 1 1.1 Control Buttons ················································································ 1 1.2 Device Symbols ················································································ 1 2. Warning, Caution and Important Tip ······························································· 2 3. Intended Use ···························································································· 2 4. Contraindications ······················································································· 3 5.
  • Page 3 BPAP System T-30T plus User Manual V2.11 17. Transferring the Device to Another Patient ······················································ 29 18. Reordering ····························································································· 30 19. Technical Support ···················································································· 30 20. Disposal ································································································ 30 21. Troubleshooting ······················································································ 31 21.1 Common Problems in Patients and Corresponding Solutions ······················· 31 21.2 Common Problems in the Device and Corresponding Solutions ···················...
  • Page 4: Symbols

    BPAP System T-30T plus User Manual V2.11 1. Symbols 1.1 Control Buttons Ramp Button Mute Button Knob 1.2 Device Symbols Follow Instructions for Use Operating Instructions Type BF Applied Part (mask) Class II (Double Insulated) AC Power DC Power ≥12.5 mm Diameter, Dripping (15ºtilted) IP22 Hot Surface No SpO...
  • Page 5: Warning, Caution And Important Tip

    BPAP System T-30T plus User Manual V2.11 2. Warning, Caution and Important Tip WARNING! Indicate the possibility of injury to the user or operator. CAUTION! Indicate the possibility of damage to the device. IMPORTANT TIP! Place emphasis on an operating characteristic. Warnings, Cautions, and Important Tips appear throughout this manual as they apply.
  • Page 6: Contraindications

    BPAP System T-30T plus User Manual V2.11 IMPORTANT! • Read and understand the entire user manual before operating this system. If you have any questions concerning the use of this system, contact your home care provider or health care professional. 4.
  • Page 7: Specifications

    BPAP System T-30T plus User Manual V2.11 5. Specifications Device Size Dimensions: 170 mm × 180 mm × 118 mm, or 290 mm × 180 mm × 134 mm (with the humidifier) Weight: 1.5 kg, or 2.5 kg (with the humidifier) Product Use, Transport and Storage Operation Transport and Storage...
  • Page 8 BPAP System T-30T plus User Manual V2.11 Static Pressure Stability ± 0.5 hPa Ramp The ramp time ranges from 0 to 60 minutes. Default setting is Off. Setting range is 150 ~ 1500 ml. Sound Pressure Level <30 dB, when the device is working at the pressure of 10 hPa. Sound Power Level <38 dB, when the device is working at the pressure of 10 hPa.
  • Page 9: Available Therapies

    BPAP System T-30T plus User Manual V2.11 6. Available Therapies The device delivers the following therapies: CPAP – Delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of pressure throughout the breathing cycle. If your health care professional has prescribed ramp for you, you can press the Ramp Button to reduce the pressure and then gradually increase the pressure to the therapeutic pressure setting so that you can fall asleep...
  • Page 10: Glossary

    BPAP System T-30T plus User Manual V2.11 7. Glossary Apnea A condition marked by the cessation of spontaneous breathing. Auto-CPAP Adjust CPAP pressure automatically to improve patient comfort based on monitoring of apnea and snoring events. Auto Off When this feature is enabled, the device automatically discontinues therapy whenever the mask is removed.
  • Page 11: Model

    BPAP System T-30T plus User Manual V2.11 Res Rate Respiratory Rate. Number of breaths per minute. Reslex A therapy feature that is enabled by your home care provider to provide pressure relief during exhalation. Standby State The state of the device when power is applied but the airflow is turned off. Means the time unit “minute”.
  • Page 12: Package Contents

    BPAP System T-30T plus User Manual V2.11 9. Package Contents After unpacking the system, make sure you have everything shown here (Different models of the product contain different components): Articles Qty. Notes Main Device Heated Humidifier Optional Shield Air Filter Power Adapter Power Cord Optional...
  • Page 13: System Features

    BPAP System T-30T plus User Manual V2.11 10. System Features Humidifier Indicator Display Screen Mute Button Power Indicator Knob Air Outlet Ramp Button Humidifier Connector Shield Fig. 10-1 Name Function Indicate the humidity level. There are five levels in total. The number of blue indicator lights that light up is directly proportional Humidifier Indicator to the humidity level.
  • Page 14: First Time Setup

    BPAP System T-30T plus User Manual V2.11 SD Card Slot Communications Port DC Inlet Filter Cap Fig. 10-2 Name Function SD Card Slot Insert the SD card into this slot Connected to external equipment (Not for connection to Communications Port unrecommended devices) DC Inlet An inlet for the DC power supply...
  • Page 15: Installing The Air Filter And Filter Cap

    BPAP System T-30T plus User Manual V2.11 the device. • Air must flow freely around the device for it to work properly. 11.2 Installing the Air Filter and Filter Cap (1) Attach the air filter to the filter cap, as shown in Fig. 11-1. Air Filter Filter Cap Fig.
  • Page 16: Assembling The Tube And Mask

    BPAP System T-30T plus User Manual V2.11 WARNINGS! • The device is powered on for use when the power cord and power adapter is connected. The Knob turns the blower On / Off. • Use of the device at an AC voltage beyond the stated range (see Section 5 “AC Power Consumption”) may damage the device or cause device failure.
  • Page 17: Using Oxygen With The Device

    BPAP System T-30T plus User Manual V2.11 port. Position the exhalation port so that the vented air is blowing away from your face. Connect the mask’s connector to the exhalation port. • If you are using a full-face mask (a mask covering both your mouth and nose), the mask must be equipped with a safety (entrainment) valve.
  • Page 18: Using The Spo 2 Kit

    BPAP System T-30T plus User Manual V2.11 SD Card Slot Fig. 11-6 If the SD card is inserted correctly, a symbol indicating correct insertion will appear in the Main Interface on the screen of the device, as shown in Fig. 11-7. Fig.
  • Page 19: Connecting The Spo 2 Kit To The Main Device

    BPAP System T-30T plus User Manual V2.11 Connector Adapter Probe Fig. 11-9 11.7.1 Connecting the SpO Kit to the Main Device (1) Pull the Adhesive-backed Paper off the Base Plate as indicated by the arrow icon in the top left of Fig. 11-10. Base Plate Adhesive-backed Paper Fig.
  • Page 20 BPAP System T-30T plus User Manual V2.11 Hook Fig. 11-13 The SpO Kit is intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO ) and pulse rate monitoring for adults weighting greater than 40 kg. It is 1.8 m in length without the function of preventing electromagnetic interference.
  • Page 21: Using The H60 Heated Humidifier

    BPAP System T-30T plus User Manual V2.11 extreme finger sizes or improper placement may degrade the SpO Kit's performance or affect the accuracy of the measurement. • Nail polish or false nails should be removed before the finger sensor is used, or it may cause erroneous measurements results.
  • Page 22: Adjusting The Tube

    BPAP System T-30T plus User Manual V2.11 CAUTION! • Before each use, examine the tube for any damage or debris. If necessary, clean the tube to remove the debris. Replace any damaged tube. Make sure that the mask does not leak. 12.2 Adjusting the Tube Lie down on your bed, and adjust the tube so it is free to move if you turn during sleep.
  • Page 23: Navigating The Patient Menu

    BPAP System T-30T plus User Manual V2.11 13. Navigating the Patient Menu 13.1 Steps to Navigating the Patient Menu 13.1.1 Accessing the Main Interface Connect the power cord and power adapter properly. The screen displays the Main Interface shown in Fig. 13-1, or the Main Interface shown in Fig. 13-2. Fig.
  • Page 24: Selecting Options

    BPAP System T-30T plus User Manual V2.11 Interface by pressing the Knob . The first option on the Setup Interface is then displayed in blue, as shown in Fig. 13-4. Fig. 13-4 13.1.4 Selecting Options As you turn the Knob clockwise, the cursor moves downwards from one option to another.
  • Page 25: Turning Pages

    BPAP System T-30T plus User Manual V2.11 Fig. 13-7 13.1.7 Turning Pages When the cursor is on Mask Type, the last option shown in Fig. 13-7, the remaining options will appear on a new page if you continue to turn the Knob clockwise, as shown in Fig.
  • Page 26 BPAP System T-30T plus User Manual V2.11 Fig. 13-10 (2) Returning to the Main Interface Move the cursor to the Home option by turning the Knob , as shown in Fig. 13-11. Fig. 13-11 Press the Knob to exit the Patient Menu. The screen will display the Main Interface shown in Fig.
  • Page 27: Options Of The Patient Menu And Corresponding Descriptions

    BPAP System T-30T plus User Manual V2.11 13.2 Options of the Patient Menu and Corresponding Descriptions Option Range Description There are five humidity levels available. As the numbering Off, 1 ~ 5 Humidifier increases, the humidity rises accordingly. “Off” means the humidifier is turned off.
  • Page 28: Alert

    BPAP System T-30T plus User Manual V2.11 0 ~ Use Time displays how long has the device been used by Use Time the user. The use time can be erased 50000 h Displays related information of the device (Model, SN, About ——...
  • Page 29 BPAP System T-30T plus User Manual V2.11 When the airflow is on, an audible alert will sound if the airway pressure is below the warning limen; the screen will display “Low Pressure!!”. Low Pressure!! Note The limens for T-30T plus: Off, 3 ~ 29 hPa, the default setting is “4 hPa”...
  • Page 30: Cleaning And Disinfection

    BPAP System T-30T plus User Manual V2.11 15. Cleaning and Disinfection WARNINGS! • Regular cleaning of the device and its accessories is very important for the prevention of respiratory infections. • To avoid electric shock, always unplug the device before cleaning. •...
  • Page 31: Replacing The Air Filter

    BPAP System T-30T plus User Manual V2.11 (2) Clean the tube in warm water which contains washing liquid, and then rinse it in clean water thoroughly. (3) After cleaning, air-dry the tube in a cool, well-ventilated area, and avoid direct sunlight. It takes approximately 30 minutes to completely air-dry the tube.
  • Page 32: Traveling With The Device

    BPAP System T-30T plus User Manual V2.11 • The device shall not be serviced or maintained while in use with a patient. • Sterilization of this device and its components other than recommended is not permitted. 16. Traveling with the Device CAUTIONS! •...
  • Page 33: Reordering

    BPAP System T-30T plus User Manual V2.11 18. Reordering Contact your home care provider to order accessories or replacement filters. The device does not require routine servicing. WARNINGS! • If you notice any unexplained changes in the performance of the device, if it is making unusual or harsh sounds, if it has been dropped or mishandled, if the enclosure is broken, or if water has entered the enclosure, discontinue use.
  • Page 34: Troubleshooting

    BPAP System T-30T plus User Manual V2.11 21. Troubleshooting The table below lists common problems you may have with the device and possible solutions to those problems. If none of the corrective actions solve the problem, contact your home care provider. 21.1 Common Problems in Patients and Corresponding Solutions Problem...
  • Page 35 BPAP System T-30T plus User Manual V2.11 Contact your physician equipment supplier. The patient is allergic to the Use a mask which is not made materials of the mask with natural rubber latex. Place a lining between the skin and mask Turn the humidity setting down, or raise the room temperature.
  • Page 36: Common Problems In The Device And Corresponding Solutions

    BPAP System T-30T plus User Manual V2.11 21.2 Common Problems in the Device and Corresponding Solutions Problem Possible Cause Solution (s) Take a few deep breaths with the The Auto On / Off feature is mask on, and the device will start enabled automatically Ensure that the power cord,...
  • Page 37: Emc Requirements

    BPAP System T-30T plus User Manual V2.11 22. EMC Requirements Guidance and manufacturer's declaration - electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The user of the device should ensure that it is used in such an environment. Electromagnetic Environment - Emissions Test Compliance...
  • Page 38 BPAP System T-30T plus User Manual V2.11 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. IEC 60601 Compliance Electromagnetic...
  • Page 39 BPAP System T-30T plus User Manual V2.11 Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The user of the device should make sure that it is used in such an environment. Immunity IEC 60601 Compliance...
  • Page 40 BPAP System T-30T plus User Manual V2.11 Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
  • Page 41 BPAP System T-30T plus User Manual V2.11 Recommended separation distances between wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment.
  • Page 42: Limited Warranty

    BPAP System T-30T plus User Manual V2.11 23. Limited Warranty ECE TIBBİ CİHAZLAR VE MEDİKAL SAN.TİC. A.Ş warrants that the device shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of one (1) year for main unit and three (3) months for all accessories from the date of sale by ECE TIBBİ...