Sony LMD-2110MD Instructions For Use Manual page 30

Hide thumbs Also See for LMD-2110MD:
Table of Contents

Advertisement

Medical Specifications
Protection against electric shock:
Class I
Protection against harmful ingress of water:
Ordinary
Degree of safety in the presence of a flammable
anesthetic mixture with air or with oxygen or
nitrous oxide:
Not suitable for use in the presence of a
flammable anesthetic mixture with air or with
oxygen or nitrous oxide
Mode of operation:
Continuous
Design and specifications are subject to change without
notice.
Note
Always verify that the unit is operating properly before
use. SONY WILL NOT BE LIABLE FOR
DAMAGES OF ANY KIND INCLUDING, BUT
NOT LIMITED TO, COMPENSATION OR
REIMBURSEMENT ON ACCOUNT OF THE LOSS
OF PRESENT OR PROSPECTIVE PROFITS DUE
TO FAILURE OF THIS UNIT, EITHER DURING
THE WARRANTY PERIOD OR AFTER
EXPIRATION OF THE WARRANTY, OR FOR
ANY OTHER REASON WHATSOEVER.
Pin assignment
PARALLEL REMOTE connector
Modular connector
(8-pin)
Pin number
Functions
1
Designating LINE input signal
2
Designating HDMI input signal
3
Designating RGB/COMPONENT input signal
4
16:9
5
GND
6
4:3
7
Selecting NORMAL
8
Selecting OVER
For details on function allocations, see REMOTE menu
(page 27).
Wiring required to use the Remote Control
Connect the function you want to use with a Remote
Control to the Ground (Pin 5).
8
1
30