Quick Start Guide
5 
Rosemount pH/ORP sensor(s) product
certifications
Rev 0.5
5.1 
European directive information
A copy of the EU Declaration of Conformity can be found at the
end of the Quick Start Guide. The most recent revision of the EU
Declaration of Conformity can be found at Emerson.com.
5.2 
Ordinary location certification
As standard, the transmitter has been examined and tested to
determine that the design meets the basic electrical, mechanical,
and fire protection requirements by a Nationally Recognized Test
Laboratory (NRTL) as accredited by the Federal Occupational Safety
and Health Administration (OSHA).
5.3 
Installing equipment in North America
The US National Electrical Code
Code (CEC) permit the use of Division marked equipment in Zones
and Zone marked equipment in Divisions. The markings must be
suitable for the area classification, gas, and temperature class. This
information is clearly defined in the respective codes.
5.4 
USA
5.4.1  USA Intrinsic Safety
Certificate
Standards
Markings
Specific Conditions for Safe Use (X):
1. Sensors with Model 1700702 preamplifier:
30
FM17US0198X
FM Class 3600:1998, FM Class 3610:2010, FM Class 3611:
2004, FM Class 3810: 2005
IS/I,II,III/1/ABCDEFG/T6 Ta = –20 °C to 60 °C
I/0/AEx ia IIC/T6 Ta = –20 °C to 60 °C
NI/I/2/ABCD/T6 Ta = –20 °C to 60 °C
S/II,III/2/EFG/T6 Ta = –20 °C to 60 °C
a. Model 385+-a-b-c. Triple junction pH/ORP sensor
b. Model 389-a-b-c-d-e. pH/ORP sensor
c. Model 389VP-a-b-c-d. pH/ORP sensor
(NEC) and the Canadian Electrical
®
Rosemount pH/ORP Sensors
May 2023