Table 3. Function Keys (cont)
Selects Input mode.
Zeros the pressure module reading when in Pressure Measurement mode.
Zeros the thermocouple TC mV/°C offset when in TC Measurement mode.
Toggles between Centigrade and Fahrenheit when you are using the TC or
RTD functions.
AUTOSET runs through the setpoints you entered using the SET function.
Press  . AUTO SET POINT? appears on the display.
Enter a number between 1 and 9 that corresponds to the number of
setpoints being used. DWELL TIME 5-500? appears on the display. Dwell
time is the number of seconds between each setpoint. The output cycles
through each setpoint and then reverses the order. For example, if 5 is
entered for the number of setpoints, the Calibrator cycles through setpoints
1, 2, 3, 4, 5 and then reverses to setpoints 4, 3, 2, and 1.
Setpoints of 30 V or greater will not go to standby when you use this
feature.
Used to enter an exponent when you define a custom RTD.
Used to recall a programmed set point.
Press  . RECALL SPT # appears on the display. Enter the number of
the output setpoint that you want to use. The output will then be
programmed to the setpoint you entered.
Used to regain local control of the Calibrator. If you set the Calibrator to a
remote state using the remote commands, all the front panel keys are
locked out except the Local key. When you press the Local key, the front
panel is unlocked.
This function does not work when you set the Calibrator using the Remote
with Lockout command. In Remote with Lockout mode, ALL keys are
locked out and the Local key will not unlock the front panel
Temperature/Pressure Calibrator
Getting Acquainted with the Calibrator
Description
Caution
Note
.
7