15B MAX/17B MAX
Users Manual

Specifications

General Specifications

Maximum voltage between
any Terminal and Earth Ground..............600 V
Maximum differential voltage
between V and COM terminals ...............1000 V
Display (LCD)..................................................6000 counts, updates
Battery Type...................................................2 AA, IEC LR6
Battery Life .....................................................500 hours minimum
Temperature
Operating.....................................................0 °C to 40 °C
Storage.........................................................-30 °C to 60 °C
Relative Humidity
Operating Humidity .................................Non-condensing
Operating Humidity, 40 MΩ range.... ≤80 % RH at 10 °C to
16
3/sec
(<10 °C); ≤90 % RH at
10 °C to 30 °C;
≤75 % RH at 30 °C to
40 °C
30 °C; ≤70 % RH at
30 °C to 40 °C
Altitude
Operating..................................................2000 m
Storage......................................................12 000 m
Temperature Coefficient...........................0.1 X (specified
Fuse protection for
W
current inputs ................................................0.44 A, 1000 V, IR 10 kA
Size (HxWxL) ..................................................183 mm x 91 mm x
Weight...............................................................455 g
Ingress Protection.......................................IEC 60529: IP40
Safety
General .......................................................IEC 61010-1: Pollution
Measurement ...........................................IEC 61010-2-033:
Electromagnetic Environment ..............IEC 61326-1: Portable
accuracy) /°C (<18 °C or
>28 °C)
11 A, 1000 V, IR 20 kA
49.5 mm
Degree 2
CAT III 600V