Hanna Instruments Total Chlorine HI 711 Instructions

Hanna Instruments Total Chlorine HI 711 Instructions

Advertisement

Quick Links

HI 711
Total Chlorine
w w w . h a n n a i n s t . c o m
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at tech@hannainst.com
or view our worldwide contact list at www.hannainst.com.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 711 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six powder reagents for Total Chlorine
• 1 x 1.5V AAA Battery
• Instruction Manual
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
Range
0.00 to 3.50 ppm
Resolution
0.01 ppm
Accuracy
±0.03 ppm ±3% of reading @ 25 °C / 77 °F
Light Emitting Diode @ 525 nm
Light Source
Light Detector
Silicon Photocell
Method
Adaptation of USEPA method 330.5. The reaction
between the chlorine and DPD reagent causes a pink
tint in the sample.
0 to 50 °C (32 to 122 °F);
Environment
max 95% RH non-condensing
Battery Type
1 x 1.5V AAA
Auto-Shut off
After 2 minutes of non-use
Dimensions
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
Weight
64 g (2.25 oz.)
Functional description:
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.

Advertisement

loading

Summary of Contents for Hanna Instruments Total Chlorine HI 711

  • Page 1 Dear Customer, Functional description: Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. If you need HI 711 additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
  • Page 2 • Turn the instrument upside down and remove the battery cover with a screwdriver. packet of HI 711-25 reagent. Replace the Hanna Instruments reserves the right to modify the design, construction or cap and shake gently for 20 seconds. Place appearance of its products without advance notice.