Specifications

INTERNAL CHANNELS

Temperature
-20 °C to +60 °C (-4 °F to +140 °F)
Range
Temperature
0.01 °C (0.018 °F)
Resolution
Calibrated
±0.5 °C (0 °C to 50 °C)
Accuracy

REMOTE CHANNELS

Remote Channel
J, K, T, E, R, S, B, N
Thermocouple
Types
Thermocouple
Female subminiature (SMP)
Connection
Cold Junction
Automatic, based on internal channel
Compensation
Thermocouple
Range (°C)
J
-210 to
+760
K
-270 to
+1370
T
-270 to
+400
E
-270 to
+980
R
-50 to
+1760
S
-50 to
+1760
B
+50 to
+1820
N
-270 to
+1300
*Thermocouple accuracy is specified with a 24 AWG.

MAXIMUM EXPOSURE TIME

Ambient
Maximum Duration in Minutes
Temperature
100 °C (212 °F)
110
150 °C (302 °F)
62
200 °C (392 °F)
45
250 °C (482 °F)
35
260 °C (500 °F)
33
300 °C (572 °F)
30
350 °C (662 °F)
25
BATTERY MAY LEAK, FLAME OR EXPLODE IF DISASSEMBLED, SHORTED, CHARGED, CONNECTED TOGETHER, MIXED WITH USED OR
OTHER BATTERIES, EXPOSED TO FIRE OR HIGH TEMPERATURE. DISCARD USED BATTERY PROMPTLY. KEEP OUT OF REACH OF CHILDREN.
6
Resolution
Accuracy*
0.1 °C
±0.5 °C
0.1 °C
±0.5 °C
0.1 °C
±0.5 °C
0.1 °C
±0.5 °C
0.5 °C
±2.0 °C
0.5 °C
±2.0 °C
0.5 °C
±2.0 °C
0.1 °C
±0.5 °C
Battery Warning
Specifications subject to change.
See Omega's terms and conditions at www.omega.com
GENERAL
4-channel: 524,032 readings per channel
Memory
8-channel: 262,016 readings per channel
(All channels
enabled)
Start Modes
Software programmable immediate start
or delay start, up to 6 months in advance
LEDs
1 per channel and 2 status LEDs
Reading Rate
4 Hz up to 1 reading every 24 hours
Calibration
Digital calibration through software
Calibration Date
Automatically recorded within device
Battery Type
9 V lithium battery included,
user replaceable
Battery Life
18 months typical
Data Format
Date and time stamped °C, °F, K, °R, mV, V
Time Accuracy
±1 minute/month
Computer
USB-A to micro USB cable (included);
Interface
460,800 baud
Operating
Windows XP SP3/7/8/10
System
Compatibility
Software
Standard Software version 4.2.17.0 or later
Compatibility
Secure Software version 4.2.16.0 or later
Operating
See table for temperature
Environment
0 %RH to 95 %RH non-condensing
Internal OM-CP-
-20 °C to +60 °C
TCTEMPX
0 %RH to 95 %RH non-condensing
OM-CP-
-20 °C to +350 °C
THERMOVAULTX
0 %RH to 95 %RH non-condensing
SYSTEM
Dimensions
12.3 in x 6.65 in x 2.45 in
(723 mm x 168 mm x 62 mm)
[Includes handles, brackets and front
enclosure latches]
9.1 in x 5.9 in x 2.45 in
(231 mm x 149 mm x 62 mm)
[Enclosure case only. Excludes handles,
brackets and latches]
Weight
4-channel: 8.85 lb (4.01 kg)
8-channel: 8.98 lb (4.07 kg)

Advertisement