Gamma Adjust - NEC MD210C3 Installation & Maintenance Manual

Medical display
Table of Contents

Advertisement

Gamma Adjust

By using the integrated front sensor, Gamma Adjust recalibrates the LCD gamma. Gamma Adjust is recommended if MAX
GSDF ERROR exceeds the medical standards. All PICTURE MODES are updated when Gamma Adjust is performed.
1. To show the CALIBRATION menu, select the STAND ALONE CALIBRATION in the advanced menu (page 29).
2. Select GAMMA ADJUST in the MODE selection (Figure 1).
3. Press SELECT to start the calibration. This may take several minutes depending on user setting.
Figure 1
4. DICOM Measurement is continuously executed automatically.
After completing DICOM Measurement, you can set DATE and TIME using control keys.
English-32