HP 8340B Manual page 36

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

4-5.
MAXIMUM
LEVELED OUTPUT POWER AND POWER ACCURACY
TEST (Cont'd)
If
the
ALC accuracy
is
tested, frequency response must
be
considered; therefore,
use the
0
dBm
power meter
indication
as
a
reference.
Also
it
should be noted that the
power
level
in
the
ENTRY DISPLAY
is
the requested
power
level.
In
normal ALC modes, the instrument
processor duplicates
the
ENTRY DISPLAY power
level
in
the
POWER
dBm
display with
0.1
dB
resolution; however,
if
AM
is
selected, the processor reads
the ALC
level
with the
A/D
converter, calculates the associated power
level, and displays the
calculated value
in
the
POWER dBm
display.
10. Press
[INSTR PRESET].
Press
[PEAK]
on.
Press[STOP FREQUENCY]
and
enter [2] [.] [3] [GHz].
Press [POWER LEVEL] and enter [0] [dBm].
11.
Press the [SWEEP
TIME]
key and enter
[2] [sec].
12.
Adjust the oscilloscope controls to view
the Power
Meter RECORDER OUTPUT voltage versus
the
DUT's SWEEP OUTPUT
voltage (A versus
B).
The oscilloscope
vertical gain and position
must
be changed as
a
function
of the power meter range and RECORDER
OUTPUT
voltage.
13. Press
Frequency Marker
key
[M2]
and, using the rotary knob, change
the
marker frequency
to
position the (intensified)
marker
on
the lowest
point on the oscilloscope trace.
It
may be necessary
to adjust the oscilloscope INTENSITY
to
view
the marker dot.
14.
Note
the
marker frequency. Press
the
[MANUAL]
key and enter the frequency noted for M2.
Reset the power meter
calibration factor switch to include
this
frequency.
The power
meter
indication
is
the minium power
point. Record
the power
meter
indication on the
work sheet
pro¬
vided
in
Table
4-10.
NOTE
For Band
0
(10
MHz
to 2.3 GHz),
if
the minimum or maximum
power
occurred below 50 MHz, use the
HP 8481A
Power Sensor
to
measure
the
power level.
15. Press
the
[CONT]
key to return to the
sweep mode.
Press
[M1]
and, using the rotary knob, change
the
marker frequency
to position the marker on the highest point on the oscilloscope trace, Note
the marker frequency. Press the
[MANUAL]
key and
enter the frequency noted for
M1.
Reset the
power meter calibration factor
switch to include
this
frequency.
The power meter
indication
is
the
maximum power
point.
Record the power meter indication
on the
work
sheet.
16.
Press the
[CONT]
key to return to
the swept mode. Press the
(START
FREQ]
key and
enter the
start frequency for the next frequency band. Press the [STOP
FREQ]
key and
enter the stop
frequency
for the next
frequency
band.
17.
Repeat steps
12
through
16
to
measure the
flatness
until all
frequency bands have been checked.
18.
The
maximum minus the minimum power meter
indications should be within
the flatness
specifi¬
cations.
If
the
DUT
does not meet
its
flatness specification, refer to
Flatness Adjustments
in
Section
V
of the Service
Manual.
ACCURACY
19.
The
absolute power
level at the
maximum and minimum
points should be within the accuracy
specification. This test may be repeated at other ALC power levels
(+10
dBm to —9.95 dBm)
to
verify flatness and accuracy specifications
over
the
ALC
range.
4-30
Performance Tests
HP 8340B/41B

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents