Hach Sension6 Quick Reference Card

Hach Sension6 Quick Reference Card

Do meter
Hide thumbs Also See for Sension6:

Advertisement

Quick Links

Setup
How To Get There
1
Turn the meter on and press
SETUP/CE. Press READ/ENTER.
2
Turn the meter on and press
SETUP/CE
Press the up arrow
.
once.
3
Turn the meter on and press
SETUP/CE
Press the up arrow
.
twice.
4
Turn the meter on and press
SETUP/CE
Press the up arrow
.
three times. Press READ/ENTER.
5
Turn the meter on and press
SETUP/CE
Press the up arrow
.
four times. Press READ/ENTER.
6
Turn the meter on and press
SETUP/CE
Press the up arrow
.
five times. Press READ/ENTER.
7
Turn the meter on and press
SETUP/CE
Press the up arrow six
.
times. Press READ/ENTER.
8
Turn the meter on and press
SETUP/CE
Press the up arrow
.
seven times. Press READ/ENTER.
9
Turn the meter on and press
SETUP/CE
Press the up arrow
.
eight times. Press READ/ENTER.
10
Turn the meter on and press
SETUP/CE
Press the up arrow
.
nine times. Press READ/ENTER.
Enter a new barometric pressure when the barometric pressure or the instrument altitude changes.
Using Sea Level Equivalent
1.
Obtain the sea level equivalent barometric
pressure from TV, radio, or a local airport.
2.
Enter this value into the meter in Setup 5.
3.
Enter the local altitude in Setup 6.
DO METER QUICK REFERENCE CARD
Table 1 Changing Meter Setup Options
Display continuous dissolved oxygen readings or lock the
reading when the value stabilizes. READ/ENTER toggles
between the options. Press EXIT when selection is
complete.
Use Metric (°C, mm Hg, meters) or English (°F, in. Hg, feet)
units by showing °C or °F. READ/ENTER toggles between
the options. Press EXIT when selection is complete.
Display D.O. concentrations in
resolution. READ/ENTER toggles between the options.
Press EXIT when selection is complete.
Adjust the displayed D.O. concentration based on sample
salinity. Use the number keys to enter a salinity value
ranging from 0–42. Press READ/ENTER and then EXIT
when selection is complete.
Use the number keys to enter the barometric pressure.
Press READ/ENTER and then EXIT when selection is
complete. See Table 2 below.
Use the number keys to enter the altitude.
Press READ/ENTER and then EXIT when selection is
complete. See Table 2 below.
Use the number keys to enter the time.
Press READ/ENTER and then EXIT when selection is
complete.
Use the number keys to enter the date.
Press READ/ENTER and then EXIT when selection is
complete.
Use the number keys to enter the year.
Press READ/ENTER and then EXIT when selection is
complete.
Use the arrow keys to select time intervals for automatic
data transfer. Press READ/ENTER and then EXIT when
selection is complete.
Table 2 Adjusting Barometric Pressure and Altitude
Description of Option
0.0 mg/L
Using True Barometric Pressure
1.
Obtain the true barometric pressure from a
nearby mercury barometer or use Table 11 on
page 74 of the manual.
2.
Enter this value into the meter in Setup 5 or
during calibration.
3.
Enter the altitude as 0 feet or meters.
51850-23
0.00 mg/L
or

Advertisement

loading

Summary of Contents for Hach Sension6

  • Page 1 51850-23 DO METER QUICK REFERENCE CARD Table 1 Changing Meter Setup Options Setup How To Get There Description of Option Turn the meter on and press Display continuous dissolved oxygen readings or lock the SETUP/CE. Press READ/ENTER. reading when the value stabilizes. READ/ENTER toggles between the options.
  • Page 2 9. When the calibration is complete, the meter will return to the read mode. Press the EXIT during the calibration sequence to back out of the calibration routine, one screen one-at-a-time, without completing a calibration. © Hach Company, 1999. All rights reserved. Printed in the U.S.A ap/dk 5/99...