Nebulizer adjustment fixture

Advertisement

Quick Links

Nebulizer Adjustment Fixture
User Guide
Before you begin
To inspect the nebulizer
To adjust or replace the nebulizer needle
Troubleshooting
This guide applies to the Nebulizer Adjustment Fixture Kit (p/n G1960-67470).
Use this kit to adjust these nebulizers:
Microflow Nebulizer Assembly (G1946-67260)
Tested ESI Nebulizer Assembly (G1946-67098)
2
3
12
ES Tested Nebulizer (G1958-67098)
Tested APCI Nebulizer Assembly (G1946-67037)
6

Advertisement

loading

Summary of Contents for Agilent Technologies G1960-67470

  • Page 1 Before you begin To inspect the nebulizer To adjust or replace the nebulizer needle Troubleshooting This guide applies to the Nebulizer Adjustment Fixture Kit (p/n G1960-67470). Use this kit to adjust these nebulizers: ES Tested Nebulizer (G1958-67098) Microflow Nebulizer Assembly (G1946-67260)
  • Page 2: Before You Begin

    Before you begin Before you begin Make sure that your kit includes these parts: Callout Number Description Part Number Nebulizer Adjustment Fixture GT430-20470 LED Flashlight 2140-0826 40× Magnifier 1000-1541 User Guide G1960-90470 ❷ ❹ ❶ ❸ Nebulizer Adjustment Fixture User Guide...
  • Page 3 To inspect the nebulizer To inspect the nebulizer 1 Insert the nebulizer into the Nebulizer Adjustment Fixture: a Carefully insert the tip of the nebulizer into the end of the Nebulizer Adjustment Fixture. b Slide the heads of the two spring loaded socket head cap screws through the mounting holes on the body of the nebulizer.
  • Page 4 To inspect the nebulizer 3 To inspect the side of the nebulizer for the correct needle adjustment length, damaged needle, or corroded nozzle (Figure a Gently push the LED Flashlight into the end horizontal position of the Nebulizer Adjustment Fixture. b Push the 40×...
  • Page 5 To inspect the nebulizer 5 Gently slide and rotate the 40× Magnifier into or out of the Nebulizer Adjustment Fixture to focus the magnifier on the end of the nebulizer nozzle and needle. For the APCI nebulizer, if needed, remove the pocket clip to focus properly.
  • Page 6 To adjust or replace the nebulizer needle To adjust or replace the nebulizer needle • When the needle is plugged. Common symptoms of a plugged needle are: When needed • increased LC back pressure • off-axis spraying • dripping from the nebulizer •...
  • Page 7 To adjust or replace the nebulizer needle Parts needed nebulizer needle kit. See Table 1. Each kit contains: • needle • ferrule • needle holder Table 1 Needle replacement kits Nebulizer Part Number ESI Nebulizer Replacement Kit used with ES Nebulizer G1958-60137 ESI Nebulizer Accessory Ship Kit used with ESI Nebulizer G1946-68703...
  • Page 8 To adjust or replace the nebulizer needle Nebulizer with locknut (original needle holder). Nebulizer with 1.5-mm hex screw. Nebulizer with T6 Torx screw (screw p/n 0515-0677) Figure 4 Nebulizer needle holders 3 Install the nebulizer in the Nebulizer Adjustment Fixture. 4 Loosen the locknut next to the zero-dead-volume (ZDV) union.
  • Page 9 To adjust or replace the nebulizer needle 7 If you need to replace the needle: a Loosen the needle holder and pull it out of the nebulizer body. Dispose of the used needle properly. b Rinse the inside of the nebulizer body with water and an organic solvent (methanol is preferred) to remove any contamination inside the nebulizer body.
  • Page 10 To adjust or replace the nebulizer needle If the needle is pushed in too far, when the union is installed and the ferrule is compressed, the HPLC flow is restricted or blocked, which results in poor or no spray from the nebulizer. h Reinstall the locknut and the union.
  • Page 11 To adjust or replace the nebulizer needle Figure 7 Side view of a correctly adjusted Tested ESI Nebulizer, with 0.003-inch dis- tance marked. Figure 8 Example of a centered needle. Nebulizer Adjustment Fixture User Guide...
  • Page 12: Troubleshooting

    Troubleshooting Troubleshooting The figures below show badly adjusted or damaged nebulizers. Problem Solution Concentricity of needle is Replace the needle. If the poor replacement needle is still out of concentricity, the nebulizer body or the needle holder is damaged or bent. Replace the complete assembly.
  • Page 13 Troubleshooting Problem Solution Needle is clogged with salt. Sonicate the nebulizer assembly in water and methanol. When sonicating, do not allow the tip of the nozzle to touch the beaker or flask. After sonication, if the needle is still clogged, replace the needle.
  • Page 14 Troubleshooting Problem Solution Needle is touching the Replace the needle. If the nozzle replacement needle is still out of concentricity, the nebulizer body is bent. Replace the nebulizer. Needle is too far out of the Adjust the needle to the nozzle.
  • Page 15 Troubleshooting Nebulizer Adjustment Fixture User Guide...
  • Page 16: In This Book

    In This Book This book contains instructions to inspect and adjust your nebulizer using the Nebulizer Adjustment Fixture.  Agilent Technologies, Inc. 2013 Printed in USA Revision A, December 2013 *G1960-90470* G1960-90470...