Replacing a Front Panel Connector

Equipment Required
Procedure
7-40
Serviceable Assemblies: Replacing a Front Panel Connector
Screwdriver
Connector (see
De-soldering tool
Soldering iron
Solder
Connector
UA (TOCO/IUP)
COMBI (US/FECG)
1.
Disconnect the AC adapter from the monitor to completely remove power.
2.
Follow the instructions for
It is not necessary to disconnect the cables. You only need access to the front
panel connectors.
3.
Follow the instructions for
no need to remove the FECG/IUP Board.)
4.
Remove the six screws from the front-end connector shield.
5.
De-solder the damaged connector and discard.
6.
Solder a new connector in place.
7.
Replace the connector shield, main board, and top cover.
170 Series Monitor
2000947-004
Table
7-14)
Table 7-14. Front Panel Connectors
Color
US/US2
Blue
White
FECG
Dark Grey
Blue and Grey
"Removing the Monitor Top Cover"
"Replacing the Main Board"
Part Number
212174
212173
212175