Thermocouple Inputs
Thermocouple
Temperature
Type
(°C)
-100.00
J
0.00
760.00
-100.00
K
0.00
1000.00
1372.00
-100.00
N
0.00
400.00
1300.00
-100.00
E
0.00
500.00
1000.00
-150.00
T
0.00
400.00
250.00
R
1000.00
1767.00
250.00
S
1000.00
1767.00
600.00
B
1000.00
1820.00
0.00
500.00
C
1000.00
1850.00
2316.00
* Sensor inaccuracies are not included.
Table 1-3. 2620A/2625A Specifications (cont)
18°C to 28°C
90 Days
1 Year
Slow
Slow
0.49
0.53
0.40
0.38
0.49
0.54
0.57
0.60
0.42
0.44
0.73
0.80
0.95
1.05
0.66
0.69
0.51
0.53
0.46
0.49
0.75
0.83
0.50
0.53
0.36
0.38
0.40
0.43
0.58
0.65
0.79
0.84
0.42
0.45
0.37
0.40
0.96
0.98
0.86
0.91
1.14
1.24
1.01
1.03
0.97
1.02
1.29
1.39
1.26
1.28
0.92
0.95
0.97
1.03
0.76
0.78
0.66
0.69
0.85
0.91
1.47
1.61
2.30
2.53
Introduction and Specifications
Accuracy (±°C)*
0°C to 60°C
1 Year
1 Year
Fast
Slow
1.00
0.73
0.77
0.53
0.97
0.91
1.20
0.82
0.88
0.57
1.46
1.36
1.89
1.85
1.48
0.90
1.14
0.66
0.99
0.72
1.53
1.45
0.99
0.75
0.72
0.52
0.77
0.71
1.11
1.16
1.66
1.16
0.89
0.58
0.74
0.61
2.48
1.14
2.10
1.29
2.65
1.96
2.62
1.20
2.37
1.42
3.02
2.17
3.52
1.40
2.48
1.16
2.41
1.51
1.87
0.92
1.53
0.96
1.90
1.41
3.18
2.70
4.93
4.35
1
Specifications
1 Year
Fast
1.22
0.91
1.35
1.43
1.02
2.03
2.70
1.70
1.29
1.23
2.16
1.22
0.86
1.05
1.63
1.99
1.04
0.97
2.65
2.48
3.38
2.80
2.77
3.80
3.64
2.69
2.89
2.01
1.81
2.41
4.29
6.77
1-9