NeoBiotech N10 Series Operation Manual

Dry bath incubator

Advertisement

Quick Links

Operation Manual
NB-12-0006
Dry Bath Incubator
N10 Series

Advertisement

Table of Contents
loading

Summary of Contents for NeoBiotech N10 Series

  • Page 1 Operation Manual NB-12-0006 Dry Bath Incubator N10 Series...
  • Page 2 Foreword Thank you for purchasing our product: Dry Bath. This manual for users contains function and operation of the instrument. In order to use the instrument properly, please read this manual carefully before using the instrument. Opening Check Please check the instrument and Appendix with the packing list when you first open the instrument packing case.
  • Page 3 Safety Warnings and Guidelines 1. Important operation information of the security: Before the users’ operation, they should have a perfect conception of how to use the Instrument. Therefore, read this Manual carefully before using it. Operation before reading the manual is forbidden. Read the guidelines and directions below and carry out the countermeasure according to them.
  • Page 4 The Instrument should be put in the place of low temperature, little dust, no water and no sun or strong lamp. What’s more, the place should be good aeration, no corrosively gas or strong disturbing magnetic field, far away from central heating, camp stove and other hot resource. Don't put the Instrument in wet and dusty place.
  • Page 5: Table Of Contents

    Contents CHAPTER 1 INTRODUCTION --------------------------------------------------------------------------------- 1 CHAPTER 2 SPECIFICATIONS ------------------------------------------------------------------------------- 2 THE NORMAL OPERATING CONDITION: ---------------------------------------------------------- 2 THE BASIC PARAMETERS AND CHARACTERISTICS ----------------------------------------- 2 OPTIONAL BLOCK ----------------------------------------------------------------------------------------- 3 CHAPTER 3 BASIC INSTRUCTIONS ----------------------------------------------------------------------- 4 STRUCTURE OVERVIEW -------------------------------------------------------------------------------- 4 KEYPADS ----------------------------------------------------------------------------------------------------- 5 DISPLAY ------------------------------------------------------------------------------------------------------- 5 CHAPTER 4 OPERATION GUIDE ---------------------------------------------------------------------------- 6 TEMPERATURE AND TIME SETTING ---------------------------------------------------------------- 6...
  • Page 6: Chapter 1 Introduction

    Dry Bath N10 Series Operation manual Chapter 1 Introduction Chapter 1 Introduction The dry bath is a new product controlled by the miniature computer, which can be widely used for the preservation and reaction of the samples, the amplification of DNA, the pre-denaturation of the electrophoresis and blood serum coagulation.
  • Page 7: Chapter 2 Specifications

    Dry Bath N10 Series Operation manual Chapter 2 Specifications Chapter 2 Specifications 1. The normal operating condition: Ambient temperature:5C  35C Relative humidity:≤70% Power supply:AC220V~ 50-60Hz 2. The basic parameters and characteristics Basic parameters: Type NB-12-0006B NB-12-0006A Parameter NB-12-0006C RT+5C  105C...
  • Page 8: Optional Block

    Dry Bath N10 Series Operation manual Chapter 2 Specifications 3. Optional block Item No. Capacity + Diameter Meas.(mm) NB-12-0006-01 96x0.2ml or 96 PCR plate 108 x 72 x 60 NB-12-0006-02 45x0.5ml Falcon 108 x 72 x 49 NB-12-0006-03 35x1.5ml Falcon...
  • Page 9: Chapter 3 Basic Instructions

    Dry Bath N10 Series Operation manual Chapter 3 Basic Instructions Chapter 3 Basic Instructions This chapter focuses on the introductions of the structure, keypads and key-functions of the instrument, as well as preparatory work before starting. Please read this chapter before starting when uses this instrument for the first time.
  • Page 10: Keypads

    Dry Bath N10 Series Operation manual Chapter 3 Basic Instructions 2. Keypads 3. Display  Drop key: For number decreasing.  Up key: For number increasing. Start For running, timing begins. Stop For stopping, timing ends. — 5 —...
  • Page 11: Chapter 4 Operation Guide

    Dry Bath N10 Series Operation manual Chapter 4 Operation Guide Chapter 4 Operation Guide 1. Temperature and time setting a) Press the On/Off button; the display flashes 3 times, the instrument goes into the initial state with the sound of “beep”.
  • Page 12: Temperature Calibration

    Dry Bath N10 Series Operation manual Chapter 4 Operation Guide thermostat countdown begins with the sound of "beep". Tips:Decimal point regularly flashing, which indicatest the instrument still in the status of temperature control, timing has not yet begin. Decimal point kept stationary, which Indicates that the instrument has had constant temperature already, the timing begins.
  • Page 13 Dry Bath N10 Series Operation manual Chapter 4 Operation Guide temperature in display is below 35℃. If the temperature is higher than 35℃, you should wait until the temperature is below 35℃. b) Inject olefin oil into one of the cone-shaped wells, and then put a thermometer into this well (make sure that the precision of the thermometer should be within 0.1℃...
  • Page 14 Dry Bath N10 Series Operation manual Chapter 4 Operation Guide If the actual read out of thermometer is 39.6℃, Modify the temperature to 39.6 by pressing  or . Then press “Start” to confirm the input value. d) Then the instrument will heat to 100℃ automatically, waiting till the instrument is in 100℃...
  • Page 15 Dry Bath N10 Series Operation manual Chapter 4 Operation Guide c) Press “Stop” key without release, then press  key at the same time for 3 seconds, it will enter into temperature calibration interface, see the right. At this point, “...
  • Page 16: 4.Temperature Control By External Sensor

    Dry Bath N10 Series Operation manual Chapter 4 Operation Guide Notes: Pressing “Stop” and “” simultaneously during the temperature calibration indicates exiting the temperature calibration Program. The changed value will be of no effect. 4.Temperature control by external sensor a) Insert the external sensor into a fixed well.
  • Page 17 Dry Bath N10 Series Operation manual Chapter 4 Operation Guide d) Screw the wrench out of the block, and fix it to another replaced block. Then put it onto the instrument and fix it well by the wrench. — 12 —...
  • Page 18: Chapter 5 Failure Analysis And Troubleshooting

    Dry Bath N10 Series Operation manual Chapter 5 Failure Analysis and Troubleshooting Chapter 5 Failure Analysis and Troubleshooting Problems and actions Common problem Possible cause Action(s) No power on the main Check power supply power plug plugged properly Faulty fuse...
  • Page 19: Annex:wiring Diagram

    Dry Bath N10 Series Operation manual Annex Annex:Wiring Diagram ( ) for reference only POWER MAIN Switch BOARD BOARD J4 Heat J2 PT100 USB1 USB J4 PT1000 HEAT PT1000 Block CONN BOARD — 14 —...