Agilent Technologies E1300B User Manual page 180

75000 series b mainframes
Hide thumbs Also See for E1300B:
Table of Contents

Advertisement

TRIGger[:IMMediate]
[:IMMediate]
Comments
Example
:SLOPe
Parameters
Comments
:SLOPe?
:SOURce
Parameters
7-52 System Instrument Command Reference
TRIGger:IMMediate will cause a trigger cycle to occur immediately, provided
that the trigger system has been initiated (INITiate).
Related Commands: ABORt, INITiate
*RST Condition: This command is an event and has no *RST condition.
Triggering the Pacer.
TRIG:SOUR HOLD
SOUR:PULS:COUN 1E3
SOUR:PULS:PER .1 S
TRIG:DELAY .75 S
INIT
TRIG
TRIGger:SLOPe < slope> is for SCPI compatibility. The mainframe's "Event
In" signal only triggers on a negative going edge.
Parameter
Parameter
Name
Type
slope
discrete
Trying to set
SLOPe to other than NEG will generate an error.
Related Commands: ABORt, INITiate,
TRIGger:SLOPe? returns the current trigger slope setting. Since the
mainframe's "Event In" signal only triggers on a negative going edge,
TRIG:SLOP? will always return "NEG".
TRIGger:SOURce < trig_source> configures the trigger system to respond to
the specified source.
Parameter
Parameter
Name
Type
trig_source
character
trigger source is TRIG
command
output 1000 Pacer pulses
pulse period set to .1 second
start Pacer .75 sec after trigger
go to Wait For Trigger state
trigger Pacer to output pulses.
Range of
Values
NEGative
Range of
Values
BUS| EXT| HOLD| IMM
Default
Units
none
Default
Units
none