BROMINE MEASUREMENT

• Fill the cuvette up to the 10 mL mark with unreacted sample and replace the cap.
• Place the cuvette into the cuvette holder and ensure that the notch on the cap is positioned
securely into the groove.
• Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
• After a few seconds, the display will show "-0.0-". The meter is now zeroed and ready for
measurement.
• Remove the cuvette.
• Add the content of one packet of
20 seconds to dissolve most of the reagent.
30
HI93716‑0
reagent. Replace the cap and shake gently for