Specifications; Available Input Ranges; Accuracy 700G Ranges; Accuracy 700Rg Ranges - Fluke FL700G01 User Manual

Pressure gauge
Table of Contents

Advertisement

Pressure Gauge

Specifications

Specifications

Available Input Ranges

See Ranges and Resolution table for available ranges in psi plus equivalent ranges and resolution for all engineering units.

Accuracy 700G Ranges

Positive Pressure ............................................................................ ±0.05 % FS
Positive Pressure (700G01, 700G02) .............................................. ±0.1 % FS
Vacuum ........................................................................................... ±0.1 % FS
Temperature Compensation ............................................................ 15 °C to 35 °C (59 °F to 95 °F) to rated accuracy
Note: For temperatures from -10 °C to 15 °C and 35 °C to 55 °C, add .003 % FS/°C

Accuracy 700RG Ranges

Positive Pressure ............................................................................ ±0.04 % rdg ±0.01 % FS
Vacuum (700RG05) ........................................................................ ±0.05 % FS
Vacuum (all other ranges) ............................................................... ±0.1 % FS
Temperature Compensation ............................................................ 0 °C to 50 °C (32 °F to 122 °F) to rated accuracy
Note
For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add .005 % FS/°C.