Dometic HC 502D Installation And Operating Manual page 17

Medication refrigerator
Hide thumbs Also See for HC 502D:
Table of Contents

Advertisement

Available languages

Available languages

HC302D, HC502D
1. Check the dimensions of the refrigerator (fig. a, page 8 or fig. b, page 8).
2. Place the refrigerator according to fig. 4, page 4 or fig. 5, page 5.
3. Mount the refrigerator to the furniture by two screws through the fittings at the
bottom (fig. 6, page 5).
4. Leave the refrigerator to stand for at least four hours before you connect it to the
power supply so that the refrigerant can flow back into the compressor.
6.4
Connecting the refrigerator to the power supply
A
NOTICE!
• Connect the refrigerator to a separate circuit to prevent
malfunctioning due to problems with other electrical equipment.
Never connect the refrigerator to a single socket with other electrical
equipment by means of a multi-socket.
• UK only: The refrigerator is supplied with a plug fitted with 3 A fuse.
In case this fuse has to be to changed, a 3 A ASTA approved
(BS 1362) fuse must be used.
1. Connect the plug of the connection cable into the AC mains.
✔ The green power LED lights up (fig. 2 2, page 3).
2. Press the
button.
✔ The display turns on.
For the first two seconds the display will not show temperature values; it shows
"--.-". Then the temperature read from the sensor is shown.
6.5
Mounting and removing the door handle
1. Mount the handle as shown (fig. 7, page 6).
2. Remove the handle as shown (fig. 8, page 6).
6.6
Installing an external alarm
Use the floating contacts at the back of the refrigerator as follows:
• The upper contacts correspond to the power fail alarm (fig. 9 1 and 2, page 7).
• The lower contacts correspond to the temperature alarm (fig. 9 3 and 4,
page 7).