Step Description
3
Note from the VAA10038.99.V00 Filter set on the
Certificate of Service Inspection for each filter:
Wavelength
Nominal value
Minimum value
Maximum value
Make sure to note the correct wavelength of the
DR300 type. Refer to
on page 43.
4
To start the service mode push and hold
at the same time.
5
Keep pushing
and release . Examine the display,
all segments must be shown. Release the key.
6
For a short time the firmware version is shown. Note
the firmware on the Certificate of Service Inspection.
7
The instrument starts the absorbance modus, the
display shows AbS.
Set the part number
, then push
Picture
Inspection
29